Multivariate standardization techniques using UV-Vis data

被引:47
作者
Sales, F
Callao, MP
Ruis, FX
机构
[1] Florenci Sales Vilalta, Departament de Química Analítica, Universitat Ravira i Virgili, 43005 Tarragona, Catalonia
关键词
UV-Vis spectroscopy; standardization; multivariate calibration;
D O I
10.1016/S0169-7439(97)00051-8
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Three standardization techniques, namely piecewise direct standardization (PDS), single wavelength standardization (SWS) and centering correction (CC) were applied to multivariate models developed using UV-Vis spectra. The data were obtained for the determination of Ca(II) and Mg(II) in natural waters analysed by sequential injection analysis (SIA). Unlike NIR, changes in measuring conditions are mainly reflected by changes in the sensitivity in the spectra. In all the models tested, the prediction errors in the best standardization conditions were less than one half of those in the non-corrected second conditions. Moreover, bias was not detected by the joint interval test. Of all the standardization techniques assayed, PDS gives the lowest mean squared error of prediction, probably due to the inclusion of more variability in the mapping of the spectra. The influential parameters, such as the number of standardization samples or the window size in PDS, were studied. The larger the number of samples in the standardization subsets, the more effective the techniques are, whereas the variation in window size does not greatly affect the calculations. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:63 / 73
页数:11
相关论文
共 31 条
  • [1] [Anonymous], 1988, J CHEMOM, DOI DOI 10.1002/CEM.1180020203
  • [2] WAVELENGTH CALIBRATION TRANSFER BETWEEN DIODE-ARRAY UV-VISIBLE SPECTROPHOTOMETERS
    BLANCO, M
    COELLO, J
    ITURRIAGA, H
    MASPOCH, S
    ROVIRA, E
    [J]. APPLIED SPECTROSCOPY, 1995, 49 (05) : 593 - 597
  • [3] Standardisation of near-infrared spectrometric instruments: A review
    Bouveresse, E
    Massart, DL
    [J]. VIBRATIONAL SPECTROSCOPY, 1996, 11 (01) : 3 - 15
  • [4] CALIBRATION TRANSFER ACROSS NEAR-INFRARED SPECTROMETRIC INSTRUMENTS USING SHENKS ALGORITHM - EFFECTS OF DIFFERENT STANDARDIZATION SAMPLES
    BOUVERESSE, E
    MASSART, DL
    DARDENNE, P
    [J]. ANALYTICA CHIMICA ACTA, 1994, 297 (03) : 405 - 416
  • [5] Standardization of near-infrared spectrometric instruments
    Bouveresse, E
    Hartmann, C
    Massart, DL
    Last, IR
    Prebble, KA
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (06) : 982 - 990
  • [6] MODIFIED ALGORITHM FOR STANDARDIZATION OF NEAR-INFRARED SPECTROMETRIC INSTRUMENTS
    BOUVERESSE, E
    MASSART, DL
    DARDENNE, P
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (08) : 1381 - 1389
  • [7] Improvement of the piecewise direct standardisation procedure for the transfer of NIR spectra for multivariate calibration
    Bouveresse, E
    Massart, DL
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 32 (02) : 201 - 213
  • [8] *CETAMA, 1986, STAT APPL EXPL MES, P213
  • [9] MULTIVARIATE CALIBRATION STANDARDIZATION
    DENOORD, OE
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1994, 25 (02) : 85 - 97
  • [10] TRANSFER OF CALIBRATION FUNCTION IN NEAR-INFRARED SPECTROSCOPY
    FORINA, M
    DRAVA, G
    ARMANINO, C
    BOGGIA, R
    LANTERI, S
    LEARDI, R
    CORTI, P
    CONTI, P
    GIANGIACOMO, R
    GALLIENA, C
    BIGONI, R
    QUARTARI, I
    SERRA, C
    FERRI, D
    LEONI, O
    LAZZERI, L
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1995, 27 (02) : 189 - 203