New mechanism to reduce the size of the monopivot magnetic suspension blood pump: Direct drive mechanism

被引:17
作者
Yamane, T [1 ]
Nishida, M [1 ]
Kijima, T [1 ]
Maekawa, J [1 ]
机构
[1] TERUMO CO LTD,CTR RES & DEV,KANAGAWA,JAPAN
关键词
centrifugal; blood pump; magnetic suspension; direct drive; totally implantable artificial heart; closed impeller;
D O I
10.1111/j.1525-1594.1997.tb03710.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Size reduction of the monopivot magnetic suspension blood pump has been achieved by reducing the size of the magnetic suspension and employing a direct drive mechanism in place of a brushless DC motor and a magnetic coupling. The flow has also been improved using a closed hollow impeller to remove now obstruction at the inlet and using radial straight vanes to reduce the impeller speed by 30%. Hemolysis testing was conducted for the new models. Results showed that model DD1 presented only a slightly higher level of hemolysis than a regular extracorporeal centrifugal pump.
引用
收藏
页码:620 / 624
页数:5
相关论文
共 4 条
[1]  
TANAKA S, 1985, JP J ARTIF ORGANS, V14, P1126
[2]   Fluid dynamics of turbo pumps for artificial hearts [J].
Yamane, T ;
Ikeda, T ;
Orita, T ;
Tateishi, T .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC MATERIALS SENSORS AND SYSTEMS, 1996, 4 (02) :99-106
[3]   DESIGN OF A CENTRIFUGAL BLOOD PUMP WITH MAGNETIC SUSPENSION [J].
YAMANE, T ;
IKEDA, T ;
ORITA, T ;
TSUTSUI, T ;
JIKUYA, T .
ARTIFICIAL ORGANS, 1995, 19 (07) :625-630
[4]  
YAMANE T, 1996, P WAS INT C MOD SIM, P42