The electrical conductivity of human cerebrospinal fluid at body temperature

被引:364
作者
Baumann, SB
Wozny, DR
Kelly, SK
Meno, FM
机构
[1] Department of Neurological Surgery, B-400, University of Pittsburgh Medical Center, University of Pittsburgh, Pittsburgh
[2] Department of Neurological Surgery, University of Pittsburgh Medical Center, University of Pittsburgh, Pittsburgh
关键词
biological tissues; brain modeling; conductivity measurements; temperature;
D O I
10.1109/10.554770
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The electrical conductivity of human cerebrospinal fluid (CSF) from seven patients was measured at both room temperature (25 degrees C) and body temperature (37 degrees C). Across the frequency range of 10Hz-10 kHz, room temperature conductivity was 1.45 S/m, but body temperature conductivity was 1.79 S/m, approximately 23% higher. Modelers of electrical sources in the human brain have underestimated human CSF conductivity by as much as 44% for nearly two decades, and this should be corrected to increase the accuracy of source localization models.
引用
收藏
页码:220 / 223
页数:4
相关论文
共 26 条
[1]   COMPLEX BIOELECTRIC IMPEDANCE MEASUREMENT SYSTEM FOR THE FREQUENCY-RANGE FROM 5 HZ TO 1 MHZ [J].
ACKMANN, JJ .
ANNALS OF BIOMEDICAL ENGINEERING, 1993, 21 (02) :135-146
[2]   A FAST METHOD FOR FORWARD COMPUTATION OF MULTIPLE-SHELL SPHERICAL HEAD MODELS [J].
BERG, P ;
SCHERG, M .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1994, 90 (01) :58-64
[3]   SPECIFIC ELECTRIC RESISTANCE OF BODY TISSUES [J].
BURGER, HC ;
VANDONGEN, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1961, 5 (04) :431-&
[4]   The electrical conductivity of animal tissues under normal and pathological conditions [J].
Crile, GW ;
Hosmer, HR ;
Rowland, AF .
AMERICAN JOURNAL OF PHYSIOLOGY, 1922, 60 (01) :59-106
[5]  
CUFFIN BN, 1979, ELECTROEN CLIN NEURO, V47, P132, DOI 10.1016/0013-4694(79)90215-3
[6]   MODELS OF THE HUMAN BRAIN AND THE SURROUNDING MEDIA - THEIR INFLUENCE ON THE RELIABILITY OF SOURCE LOCALIZATION [J].
FENDER, DH .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1991, 8 (04) :381-390
[7]  
Fishman R.A., 1992, CEREBROSPINAL FLUID, V2nd
[8]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[9]  
GEDDES LA, 1972, ELECTRODES MEASUREME, P32
[10]   REALISTIC CONDUCTIVITY GEOMETRY MODEL OF THE HUMAN HEAD FOR INTERPRETATION OF NEUROMAGNETIC DATA [J].
HAMALAINEN, MS ;
SARVAS, J .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (02) :165-171