Transvenous closure of secundum atrial septal defects - Preliminary results with a new self-expanding nitinol prosthesis in a swine model

被引:183
作者
Sharafuddin, MJA
Gu, XP
Titus, JL
Urness, M
CerveraCeballos, JJ
Amplatz, K
机构
[1] UNIV MINNESOTA HOSP & CLIN,DEPT RADIOL,SECT CARDIOVASC & INTERVENT RADIOL,MINNEAPOLIS,MN 55455
[2] UNIV MINNESOTA HOSP & CLIN,DEPT PATHOL LAB MED,MINNEAPOLIS,MN 55455
[3] UNITED HOSP,JESSE E EDWARDS REGISTRY CARDIOVASC DIS,ST PAUL,MN
关键词
heart septal defects; heart defects; congenital; pediatrics; shunts;
D O I
10.1161/01.CIR.95.8.2162
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Our purpose was to evaluate a new prosthesis for percutaneous closure of secundum atrial septal defects (ASDs). Methods and Results Percutaneous closure of surgically created fossa ovalis ASD was attempted in lj minipigs. The mean balloon-stretched ASD diameter was 12.3+/-2.3 mm (range, 10 to 16 mm). The self-expanding prosthesis was braided from 0.005-in Nitinol wires in the shape of two flat buttons with a short connecting waist with a diameter corresponding to that of the defect to be closed. Polyester filling was added to enhance thrombogenicity. Pulmonary arteriography with levo-phase was obtained before placement; immediately after placement and at 1-week, 1-month, and 3-month follow-ups. Four animals were killed at 1 week, 1 month, and 3 months for histopathological correlation. Three deaths resulted from ventricular fibrillation (one during anesthesia and two during the placement procedure). Successful placement of the prosthesis was achieved in the remaining 12 animals. Overall immediate ASD closure on angiography Occurred in 7 of 12 animals (all polyester-filled prostheses). Absent or trace shunt by angiography was present in 11 of 12 devices at 1 week, with the remaining one demonstrating a small shunt. All septal defects were completely closed at 1 month with the exception of one case in which delayed partial dislodgment of an undersized prosthesis into the right atrium had developed. Closure rate at 3 months was 100%. Neoendothelialization and fibrous incorporation of the prosthesis were completed within 1 to 3 months. Conclusions Effective and permanent occlusion of secundum ASDs is feasible with a device that offers the advantages of easy placement, self-centering, and repositionability.
引用
收藏
页码:2162 / 2168
页数:7
相关论文
共 31 条
[1]   Failure of devices used for closure of atrial septal defects: Mechanisms and management [J].
Agarwal, SK ;
Ghosh, PK ;
Mittal, PK .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 112 (01) :21-26
[2]   TRANSCATHETER CLOSURE OF PATENT FORAMEN OVALE AFTER PRESUMED PARADOXICAL EMBOLISM [J].
BRIDGES, ND ;
HELLENBRAND, W ;
LATSON, L ;
FILIANO, J ;
NEWBURGER, JW ;
LOCK, JE .
CIRCULATION, 1992, 86 (06) :1902-1908
[3]   A NEW CANINE MODEL FOR EVALUATING BLOOD PROSTHETIC ARTERIAL GRAFT INTERACTIONS [J].
BROPHY, CM ;
ITO, RK ;
QUIST, WC ;
ROSENBLATT, MS ;
CONTRERAS, M ;
TSOUKAS, A ;
LOGERFO, FW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (08) :1031-1038
[4]  
CHAN KC, 1993, BRIT HEART J, V69, P52
[5]   NITINOL INTRAVASCULAR STENT - RESULTS OF PRECLINICAL EVALUATION [J].
CRAGG, AH ;
DEJONG, SC ;
BARNHART, WH ;
LANDAS, SK ;
SMITH, TP .
RADIOLOGY, 1993, 189 (03) :775-778
[6]   EXPERIMENTAL ATRIAL SEPTAL-DEFECT CLOSURE WITH A NEW, TRANSCATHETER, SELF-CENTERING DEVICE [J].
DAS, GS ;
VOSS, G ;
JARVIS, G ;
WYCHE, K ;
GUNTHER, R ;
WILSON, RF .
CIRCULATION, 1993, 88 (04) :1754-1764
[7]   SHAPE-MEMORY ALLOYS AS NEW MATERIALS FOR ASEISMIC ISOLATION [J].
GRAESSER, EJ ;
COZZARELLI, FA .
JOURNAL OF ENGINEERING MECHANICS-ASCE, 1991, 117 (11) :2590-2608
[8]   Transcatheter closure of secundum atrial septal defects with the atrial septal defect occlusion system (ASDOS): Initial experience in children [J].
Hausdorf, G ;
Schneider, M ;
Franzbach, B ;
Kampmann, C ;
Kargus, K ;
Goeldner, B .
HEART, 1996, 75 (01) :83-88
[9]  
HEBDA DA, 1994, HYSTERESIS TESTING N, P1
[10]  
HICKEY PR, 1992, ANESTH ANALG, V74, P44