Revision strategies for single- and two-level total disc arthroplasty procedures: a biomechanical perspective

被引:20
作者
Cunningham, Bryan W. [1 ,2 ]
Hu, Nianbin [1 ,2 ]
Beatson, Helen J. [1 ,2 ]
Serhan, Hassan [3 ]
Sefter, John C. [1 ,2 ]
McAfee, Paul C. [1 ,2 ]
机构
[1] St Joseph Med Ctr, Orthopaed Spinal Res Lab, Towson, MD 21204 USA
[2] St Joseph Med Ctr, Scoliosis & Spine Ctr, Towson, MD 21204 USA
[3] DePuy Spine Inc, Raynham, MA USA
关键词
Multilevel total disc arthroplasty; Revision strategies; Motion preservation; CHARITE(TM) ARTIFICIAL DISC; HUMAN CADAVERIC MODEL; IN-VITRO; CERVICAL JOINT; LUMBAR SPINE; FOLLOW-UP; REPLACEMENT; ADJACENT; PROSTHESIS; KINEMATICS;
D O I
10.1016/j.spinee.2009.03.011
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
BACKGROUND CONTEXT: The utilization of motion-preserving implants versus conventional instrumentation systems, which stabilize the operative segments, necessitates improved understanding of their comparative biomechanical properties and optimal biomechanical method for surgical revision. PURPOSE: Using an in vitro human cadaveric model, the primary objective was to compare the multidirectional flexibility properties of single- versus two-level total disc arthroplasty procedures and determine the acute in vitro biomechanical characteristics of two methods of surgical revision-posterior transpedicular instrumentation alone or circumferential spinal arthrodesis. STUDY DESIGN: This in vitro biomechanical study was undertaken to compare the multidirectional flexibility kinematics of single- versus two-level lumbar total disc arthroplasty reconstructions using an in vitro model. METHODS: A total of seven human cadaveric lumbosacral spines (L1-sacrum) were biomechanically evaluated under the following L4-L5 reconstruction conditions: intact spine; discectomy alone; Charite total disc replacement; Charite with pedicle screws; two-level Charite (L4-S1); two-level Charite with pedicle screws (L4-S1); Charite L4-L5 with pedicle screws and femoral ring allograft (FRA) (L5-S1); and pedicle screws with FRA (L4-S1). Multidirectional flexibility testing used the Panjabi Hybrid Testing protocol, which includes pure moments for the intact condition with the overall spinal motion replicated under displacement control for subsequent reconstructions. Hence, changes in adjacent level kinematics can be obtained compared with pure moment testing strategies. Unconstrained intact moments of +/-7.5 Nm were used for axial rotation, flex ion-extension, and lateral bending testing with quantification of the operative- and adjacent-level range of motion (ROM). All data were normalized to the intact spine condition (intact=100%). RESULTS: In axial rotation, single- and two-level Charite reconstructions produced significantly more motion than pedicle screw constructs combined with the Charite or FRA (p<.05), There were no differences between the Charite augmented with pedicle screws or pedicle screws with FRA (p>.05). The two-level annulus lumbar resection required for multilevel Charite implantation had an added destabilizing effect, resulting in a 140% to 160% ROM increase over the intact condition. Under two-level reconstructions, rotational motion at the L4-L5 level increased from 160+/-26% to 263+/-65% with the implantation of the second Charite at L5-S1. Flexion-extension and lateral bending conditions with the Charite reconstructions in this group of seven spines demonstrated no significant differences compared with the intact spine (p>.05). The Charite combined with pedicle screws or pedicle screws with FRA significantly reduced motion at the operative level compared with the Charite reconstruction (p<.05). The most pronounced changes in adjacent level kinematics and intradiscal pressures were observed under flex ion-extension loading. The addition of pedicle screw fixation increased segmental motion and intradiscal pressures at the proximal and distal adjacent levels compared with the intact and Charite reconstruction groups (p<.05). CONCLUSIONS: The findings highlight a variety of important trends at the operative and adjacent levels. In terms of revision strategies, posterior pedicle screw reconstruction combined with an existing Charite was not found acutely to be statistically different from pedicle screws combined with FRA. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:735 / 743
页数:9
相关论文
共 27 条
[1]
AHRENS J, 1998, NORMAL JOINT MOBILIT, P56
[2]
The artificial disc: Theory, design and materials [J].
Bao, QB ;
McCullen, GM ;
Higham, PA ;
Dumbleton, JH ;
Yuan, HA .
BIOMATERIALS, 1996, 17 (12) :1157-1167
[3]
Biomechanics of lumbar arthroplasty [J].
Crawford, NR .
NEUROSURGERY CLINICS OF NORTH AMERICA, 2005, 16 (04) :595-+
[4]
A minimally disruptive technique for measuring intervertebral disc pressure in vitro: application to the cervical spine [J].
Cripton, PA ;
Dumas, GA ;
Nolte, LP .
JOURNAL OF BIOMECHANICS, 2001, 34 (04) :545-549
[5]
Cunningham Bryan W, 2004, Spine J, V4, p219S, DOI 10.1016/j.spinee.2004.07.015
[6]
Biomechanical evaluation of total disc replacement arthroplasty:: An in vitro human cadaveric model [J].
Cunningham, BW ;
Gordon, JD ;
Dmitriev, AE ;
Hu, NB ;
McAfee, PC .
SPINE, 2003, 28 (20) :S110-S117
[7]
Biomechanical testing of an artificial cervical joint and an anterior cervical plate [J].
DiAngelo, DJ ;
Roberston, JT ;
Metcalf, NH ;
McVay, BJ ;
Davis, RC .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2003, 16 (04) :314-323
[8]
Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty -: An In Vitro human cadaveric model [J].
Dmitriev, AE ;
Cunningham, BW ;
Hu, NB ;
Sell, G ;
Vigna, F ;
McAfee, PC .
SPINE, 2005, 30 (10) :1165-1172
[9]
Effects of Charite artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol [J].
Goel, VK ;
Grauer, JN ;
Patel, TC ;
Biyani, A ;
Sairyo, K ;
Vishnubhotla, S ;
Matyas, A ;
Cowgill, I ;
Shaw, M ;
Long, R ;
Dick, D ;
Panjabi, MM ;
Serhan, H .
SPINE, 2005, 30 (24) :2755-2764
[10]
A MULTICENTER RETROSPECTIVE STUDY OF THE CLINICAL-RESULTS OF THE LIN(R) SB CHARITE INTERVERTEBRAL PROSTHESIS - THE INITIAL EUROPEAN EXPERIENCE [J].
GRIFFITH, SL ;
SHELOKOV, AP ;
BUTTNERJANZ, K ;
LEMAIRE, JP ;
ZEEGERS, WS .
SPINE, 1994, 19 (16) :1842-1849