Pronase-immobilized enzyme reactor:: An approach for automation in glycoprotein analysis by LC/LC-ESI/MSn

被引:58
作者
Temporini, Caterina
Perani, Eleonora
Calleri, Enrica
Dolcini, Lorenzo
Lubda, Dieter
Caccialanza, Gabriele
Massolini, Gabriella
机构
[1] Univ Pavia, Dipartimento Chim Farmaceut, I-27100 Pavia, Italy
[2] Univ Pavia, Dipartimento Biochim, I-27100 Pavia, Italy
[3] Merck KGaA, LSP R&D MDA, D-64293 Darmstadt, Germany
关键词
D O I
10.1021/ac0611519
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated analytical approach is proposed for simultaneous characterization of glycan and peptide moieties in pronase-generated glycopeptides. The proposed method is based on the use of a new pronase-immobilized enzyme reactor for the on-line rapid digestion of the target glycoprotein. By coupling the bioreactor to a Hypercarb chromatographic trap column, on-line selective glycopeptide enrichment prior to normal-phase liquid chromatography-mass spectrometry was obtained. A detailed study was carried out for integration and automation of each phase of the proposed analytical procedure. On-line digestion allowed extensive cleavage of the model protein (ribonuclease B), yielding to glycopeptides with peptide moieties up to eight amino acids, carrying the Man(5)-Man(9) N-glycans each, selectively resolved on an Amide-80 column. The use of a linear ion trap instrument resulted in efficient ion capture and led to MS3 acquisition times and spectra quality similar to those for MS2, allowing the unambiguous identification of glycan (MS2) and peptide (MS3) sequences. The proposed procedure reduces the glycoprotein analysis time from similar to 3 days, as in most of the traditional off-line methods, to similar to 1 h.
引用
收藏
页码:355 / 363
页数:9
相关论文
共 45 条
[1]   Determination of N-glycosylation sites and site heterogeneity in glycoproteins [J].
An, HJ ;
Peavy, TR ;
Hedrick, JL ;
Lebrilla, CB .
ANALYTICAL CHEMISTRY, 2003, 75 (20) :5628-5637
[2]   The estimation of pepsin, trypsin, papain, and cathepsin with hemoglobin [J].
Anson, ML .
JOURNAL OF GENERAL PHYSIOLOGY, 1938, 22 (01) :79-89
[3]   Characterization of protein glycoforms by capillary-zone electrophoresis nanoelectrospray mass spectrometry [J].
Bateman, KP ;
White, RL ;
Yaguchi, M ;
Thibault, P .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :327-344
[4]   Evaluation of a monolithic epoxy silica support for penicillin G acylase immobilization [J].
Calleri, E ;
Massolini, G ;
Lubda, D ;
Temporini, C ;
Loiodice, F ;
Caccialanza, G .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1031 (1-2) :93-100
[5]   Development of a bioreactor based on trypsin immobilized on monolithic support for the on-line digestion and identification of proteins [J].
Calleri, E ;
Temporini, C ;
Perani, E ;
Stella, C ;
Rudaz, S ;
Lubda, D ;
Mellerio, G ;
Veuthey, JL ;
Caccialanza, G ;
Massolini, G .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1045 (1-2) :99-109
[6]   Trypsin-based monolithic bioreactor coupled on-line with LC/MS/MS system for protein digestion and variant identification in standard solutions and serum samples [J].
Calleri, E ;
Temporini, C ;
Perani, E ;
De Palma, A ;
Lubda, D ;
Mellerio, G ;
Sala, A ;
Galliano, M ;
Caccialanza, G ;
Massolini, G .
JOURNAL OF PROTEOME RESEARCH, 2005, 4 (02) :481-490
[7]  
Cooper CA, 2001, PROTEOMICS, V1, P340, DOI 10.1002/1615-9861(200102)1:2<340::AID-PROT340>3.3.CO
[8]  
2-2
[9]   Glycoprotein structure determination mass spectrometry [J].
Dell, A ;
Morris, HR .
SCIENCE, 2001, 291 (5512) :2351-2356
[10]   Determination of glycopeptide structures by multistage mass spectrometry with low-energy collision-induced dissociation: comparison of electrospray ionization quadrupole ion trap and matrix-assisted laser desorption/ionization quadrupole ion trap reflectron time-of-flight approaches [J].
Demelbauer, UM ;
Zehl, M ;
Plematl, A ;
Allmaier, G ;
Rizzi, A .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (14) :1575-1582