Clip placement after stereotactic vacuum-assisted breast biopsy

被引:133
作者
Liberman, L [1 ]
Dershaw, DD [1 ]
Morris, EA [1 ]
Abramson, AF [1 ]
Thornton, CM [1 ]
Rosen, PP [1 ]
机构
[1] MEM SLOAN KETTERING CANC CTR,DEPT PATHOL,NEW YORK,NY 10021
关键词
biopsies; technology; breast; biopsy; breast neoplasms; diagnosis; stereotaxis;
D O I
10.1148/radiology.205.2.9356622
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To assess accuracy and usefulness of placement of a localizing clip after stereotactic, vacuum-assisted breast biopsy. MATERIALS AND METHODS: Retrospective review was performed of 57 lesions that underwent placement of a localizing clip after stereotactic vacuum-assisted biopsy with an 11-gauge (n = 42) or 14-gauge (n = 15) probe. The clip was placed when images obtained after stereotactic biopsy suggested that the lesion seen at mammography was removed. Coordinates of the clip on stereotactic images obtained after placement were compared with lesion coordinates determined before biopsy. Surgery was performed in 25 cases. Mammographic and histopathologic findings were reviewed. RESULTS: The distance from clip to lesion site was less than 1 cm in 40 (95%) of 42 lesions that underwent clip placement with the Ii-gauge probe versus 11 (73%) of 15 lesions that underwent clip placement after 14-gauge biopsy (P < .04). The biopsy site was identified in the surgical specimen in 19 (100%) lesions with clips after Ii-gauge biopsy and five (83%) of six lesions with clips after 14-gauge biopsy. No complications occurred. CONCLUSION: A localizing clip can be placed in proximity to the stereotactic biopsy site through an Ii-gauge probe. Clip placement can enable accurate localization for surgical excision.
引用
收藏
页码:417 / 422
页数:6
相关论文
共 19 条
[11]  
KANE JW, 1978, PHYSICS, P627
[12]   Interventional breast MR imaging: Clinical use of a stereotactic localization and biopsy device [J].
Kuhl, CK ;
Elevelt, A ;
Leutner, CC ;
Gieseke, J ;
Pakos, E ;
Schild, HH .
RADIOLOGY, 1997, 204 (03) :667-675
[13]   Mammographic findings after stereotactic 14-gauge vacuum biopsy [J].
Liberman, L ;
Hann, LE ;
Dershaw, DD ;
Morris, EA ;
Abramson, AF ;
Rosen, PP .
RADIOLOGY, 1997, 203 (02) :343-347
[14]   Stereotactic breast biopsy of clustered microcalcifications with a directional, vacuum-assisted device [J].
Meyer, JE ;
Smith, DN ;
DiPiro, PJ ;
Denison, CM ;
Frenna, TH ;
Harvey, SC ;
Ko, WD .
RADIOLOGY, 1997, 204 (02) :575-576
[15]   Laser therapy for breast cancer: MR imaging and histopathologic correlation [J].
Mumtaz, H ;
HallCraggs, MA ;
Wotherspoon, A ;
Paley, M ;
Buonaccorsi, G ;
Amin, Z ;
Wilkinson, I ;
Kissin, MW ;
Davidson, I ;
Taylor, I ;
Bown, SG .
RADIOLOGY, 1996, 200 (03) :651-658
[16]   MR IMAGING-GUIDED LOCALIZATION AND BIOPSY OF BREAST-LESIONS - INITIAL EXPERIENCE [J].
OREL, SG ;
SCHNALL, MD ;
NEWMAN, RW ;
POWELL, CM ;
TOROSIAN, MH ;
ROSATO, EF .
RADIOLOGY, 1994, 193 (01) :97-102
[17]   STAGING OF SUSPECTED BREAST-CANCER - EFFECT OF MR-IMAGING AND MR-GUIDED BIOPSY [J].
OREL, SG ;
SCHNALL, MD ;
POWELL, CM ;
HOCHMAN, MG ;
SOLIN, LJ ;
FOWBLE, BL ;
TOROSIAN, MH ;
ROSATO, EF .
RADIOLOGY, 1995, 196 (01) :115-122
[18]  
PARKER SH, 1995, RADIOL CLIN N AM, V33, P1171
[19]   A practical approach to minimally invasive breast biopsy [J].
Parker, SH ;
Burbank, F .
RADIOLOGY, 1996, 200 (01) :11-20