Effect of respiratory motion on the delivery of breast radiotherapy using SMLC intensity modulation

被引:21
作者
Liu, Qiang
McDermott, Patrick
Burmeister, Jay [1 ]
机构
[1] Beth Israel Deaconess Med Ctr, Dept Radiat Therapy, Boston, MA 02215 USA
[2] William Beaumont Hosp, Royal Oak, MI 48073 USA
[3] Wayne State Univ, Karmanos Canc Inst, Gershenson Radiat Oncol Ctr, Detroit, MI 48201 USA
关键词
breast IMRT; respiratory motion; film dosimetry;
D O I
10.1118/1.2405323
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study evaluates the effects of respiratory motion on breast radiotherapy delivered using segmented multileaf collimator (SMLC) intensity modulation. An anthropomorphic breast phantom was constructed of polystyrene plates between which radiographic films were inserted. The phantom was mounted on a moving platform to simulate one-dimensional sinusoidal oscillation with variable amplitude and frequency. The motion effect on two breast IMRT techniques, a beamlet-based plan created using the Corvus treatment planning system and an aperture-based plan, was evaluated via film comparison. Motion-induced differences in the treatment region are generally within +/- 5%, with the exception of the posterior field edge and the apex of the breast in the Corvus IMRT plan. Considering the experimental uncertainty arising from the setup and film dosimetry, this result indicates that respiratory motion-induced dose variations are generally relatively insignificant. It appears that the anterior hot spots observed in the Corvus IMRT plan result from the high intensity fluence delivered to the "virtual bolus" area which must be created during the planning process in order to provide "flash" for the respiratory motion. The potential magnitude of such effects resulting from the interplay between fluence modulation and patient motion are unique to the individual planning system and planning technique, as well as the delivery equipment and technique. Such effects should be carefully investigated prior to the implementation of IMRT for breast radiotherapy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:347 / 351
页数:5
相关论文
共 12 条
[1]   Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation [J].
Bortfeld, T ;
Jokivarsi, K ;
Goitein, M ;
Kung, J ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) :2203-2220
[2]   Effect of MLC leaf width on the planning and delivery of SMLC IMRT using the CORVUS inverse treatment planning system [J].
Burmeister, J ;
McDermott, PN ;
Bossenberger, T ;
Ben-Josef, E ;
Levin, K ;
Forman, JD .
MEDICAL PHYSICS, 2004, 31 (12) :3187-3193
[3]   The effects of intra-fraction organ motion on the delivery of intensity-modulated field with a multileaf collimator [J].
Chui, CS ;
Yorke, E ;
Hong, L .
MEDICAL PHYSICS, 2003, 30 (07) :1736-1746
[4]   The effect of respiratory motion on CSI localized MRS [J].
Doyle, VL ;
Howe, FA ;
Griffiths, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2093-2104
[5]   Impact of breathing motion on whole breast radiotherapy: A dosimetric analysis using active breathing control [J].
Frazier, RC ;
Vicini, FA ;
Sharpe, MB ;
Yan, D ;
Fayad, J ;
Baglan, KL ;
Kestin, LL ;
Remouchamps, VM ;
Martinez, AA ;
Wong, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (04) :1041-1047
[6]   Quantifying the effect of intrafraction motion during breast IMRT planning and dose delivery [J].
George, R ;
Keall, PJ ;
Kini, VR ;
Vedam, SS ;
Siebers, JV ;
Wu, Q ;
Lauterbach, MH ;
Arthur, DW ;
Mohan, R .
MEDICAL PHYSICS, 2003, 30 (04) :552-562
[7]   An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments [J].
Jiang, SB ;
Pope, C ;
Al Jarrah, KM ;
Kung, JH ;
Bortfeld, T ;
Chen, GTY .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (12) :1773-1784
[8]   Respiration gated radiotherapy treatment: A technical study [J].
Kubo, HD ;
Hill, BC .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :83-91
[9]  
LIU Q, 2004, THESIS WAYNE STATE U
[10]   Influence of intra-fractional breathing movement in step-and-shoot IMRT [J].
Schaefer, M ;
Münter, MW ;
Thilmann, C ;
Sterzing, F ;
Haering, P ;
Combs, SE ;
Debus, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :N175-N179