共 13 条
Determination of cysteine by capillary zone electrophoresis with end-column amperometric detection at a gold/mercury amalgam microelectrode without deoxygenation
被引:109
作者:
Jin, WR
Wang, Y
机构:
[1] Laboratory of Analytical Science, Sch. of Chem., Shandong University
关键词:
electrochemical detection;
detection;
electrophoresis;
electrodes;
cysteine;
amino acids;
D O I:
10.1016/S0021-9673(97)00015-0
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
Capillary zone electrophoresis was employed for the determination of cysteine using an end-column amperometric detection with a gold/mercury amalgam microeIectrode, at a constant potential of 0.02-0.06 V vs. saturated calomel electrode. In this procedure deoxygenation is not necessary. The electrochemical characteristics at the microelectrode, the effect of the concentration of the buffer and the separation voltage across the capillary on the migration time and separation efficiency, and the dependence of the injection voltage and time on the detection signal, the separation efficiency and coulometric efficiency has been investigated. The calibration plot was found to be linear over four orders of magnitude and the limit of detection was 5.8 . 10(-8) mol/l (or 14.5 amol). The method was applied to the determination of cysteine in human plasma, blood and urine.
引用
收藏
页码:307 / 314
页数:8
相关论文