Use of region II of the a/q stability diagram for fast scanning of a linear quadrupole mass spectrometer

被引:13
作者
Grimm, CC [1 ]
Clawson, R [1 ]
Short, RT [1 ]
机构
[1] OAK RIDGE NATL LAB,OAK RIDGE,TN
关键词
FILTER;
D O I
10.1016/S1044-0305(97)00006-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
For evaluation as a detector for high-speed gas chromatography, a linear quadrupole mass spectrometer was operated in rf only mode by using Region II (a = 0; 7.514 < q < 7.580) of the Mathieu a/q stability diagram. The available power supply and the diameter of the quadrupole rods of the mass spectrometer placed an upper mass limit of similar to m/z 93. Scan rates of 1000 scans/s were obtained with mass spectral peaks resolved over an 80-u range. The m/z 91 and 92 ions produced from the electron ionization of toluene are resolved with an R-1/2 of 135. A potential difference between the source and the quadruple mass filter of up to 1000 V was used to accelerate ions into the quadrupole. Broadening of mass-to-charge ratio peaks results from the time constant of the signal amplification rather than the small number of rf cycles the ions experience. The expected loss of sensitivity relative to Region I is observed, and the problem of mass aliasing is discussed. (C) 1997 American Society for Mass Spectrometry.
引用
收藏
页码:539 / 544
页数:6
相关论文
共 16 条
[1]  
AMIRAV A, 1995, J AM SOC MASS SPECTR, V6, P120
[2]   ORTHOGONAL ACCELERATION - A NEW DIRECTION FOR TIME-OF-FLIGHT MASS-SPECTROMETRY - FAST, SENSITIVE MASS ANALYSIS FOR CONTINUOUS ION SOURCES [J].
COLES, J ;
GUILHAUS, M .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1993, 12 (05) :203-213
[3]  
Dagan S, 1996, J AM SOC MASS SPECTR, V7, P737, DOI 10.1016/1044-0305(96)80519-8
[4]  
Dawson P., 1995, AVS CLASSICS VACUUM
[5]   THE 2ND STABILITY REGION OF THE QUADRUPOLE MASS FILTER .1. ION OPTICAL-PROPERTIES [J].
DAWSON, PH ;
YU, BQ .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1984, 56 (01) :25-39
[6]   THE 2ND STABILITY REGION OF THE QUADRUPOLE MASS FILTER .2. EXPERIMENTAL RESULTS [J].
DAWSON, PH ;
YU, BQ .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1984, 56 (01) :41-50
[7]  
GRIMM CC, 1996, AM LAB, V7, pS18
[8]   APPLICABILITY OF THE RF-ONLY MODE IN A QUADRUPOLE MASS-SPECTROMETER FOR DIRECT ANALYSIS OF KEV IONS [J].
HIROKI, S ;
KANEKO, K ;
MURAKAMI, Y .
VACUUM, 1995, 46 (12) :1445-1447
[9]   MASS-SPECTROMETRY ON THE CHROMATOGRAPHIC TIME SCALE - REALISTIC EXPECTATIONS [J].
HOLLAND, JF ;
ENKE, CG ;
ALLISON, J ;
STULTS, JT ;
PINKSTON, JD ;
NEWCOME, B ;
WATSON, JT .
ANALYTICAL CHEMISTRY, 1983, 55 (09) :A997-&
[10]  
HOLLAND JF, 1994, ACS SYM SER, V549, P157