Image reconstruction for photoacoustic scanning of tissue structures

被引:201
作者
Hoelen, CGA [1 ]
de Mul, FFM [1 ]
机构
[1] Univ Twente, Dept Appl Phys, NL-7500 AE Enschede, Netherlands
关键词
D O I
10.1364/AO.39.005872
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Photoacoustic signal generation can be used for a new medical tomographic technique. This makes it possible to image optically different structures, such as the (micro)vascular system in tissues, by use of a transducer array for the detection of laser-generated wide-bandwidth ultrasound. A time-domain delay-and-sum focused beam-forming technique is used to locate the photoacoustic sources in the sample. To characterize the transducer response, simulations have been performed for a wide variety of parameter values and have been verified experimentally. With these data the weight factors for the spectrally and temporally filtered sensor signals are determined in order to optimize the signal-to-noise ratio of the beam former. The imaging algorithm is investigated by simulations and experiments. With this algorithm, for what is to our knowledge the first time, the three-dimensional photoacoustic imaging of complex optically absorbing structures located in a highly diffuse medium is demonstrated. When 200-mum-diameter hydrophone elements are used, the depth resolution is better than 20 mum, and the lateral resolution is better than 200 mum, independent of the depth for our range of imaging (to similar to6 mm). Reduction of the transducer diameters and adaptation of the weight factors, at the cost of some increase of the noise level, will further improve the lateral resolution. The synthetic aperture algorithm used has been shown to be suitable for the new technique of photoacoustic tissue scanning. (C) 2000 Optical Society of America OCIS codes: 110.5120, 170.3880.
引用
收藏
页码:5872 / 5883
页数:12
相关论文
共 27 条
[1]  
DIEBOLD GJ, 1994, ACUSTICA, V80, P339
[2]   FUNDAMENTALS OF DIGITAL ARRAY PROCESSING [J].
DUDGEON, DE .
PROCEEDINGS OF THE IEEE, 1977, 65 (06) :898-904
[3]   Laser optoacoustic imaging for breast cancer diagnostics: Limit of detection and comparison with X-ray and ultrasound imaging [J].
Esenaliev, RO ;
Karabutov, AA ;
Tittel, FK ;
Fornage, BD ;
Thomsen, SL ;
Stelling, C ;
Oraevsky, AA .
OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE: THEORY, INSTRUMENTATION, MODEL, AND HUMAN STUDIES II, PROCEEDINGS OF, 1997, 2979 :71-82
[4]   Sensitivity of laser opto-acoustic imaging in detection of small deeply embedded tumors [J].
Esenaliev, RO ;
Karabutov, AA ;
Oraevsky, AA .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 1999, 5 (04) :981-988
[5]  
EVANS JA, 1988, PRACTICAL ULTRASOUND, P15
[6]  
EVANS JA, 1988, PRACTICAL ULTRASOUND, P1
[7]   A new theoretical approach to photoacoustic signal generation [J].
Hoelen, CGA ;
de Mul, FFM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (02) :695-706
[8]   Three-dimensional photoacoustic imaging of blood vessels in tissue [J].
Hoelen, CGA ;
de Mul, FFM ;
Pongers, R ;
Dekker, A .
OPTICS LETTERS, 1998, 23 (08) :648-650
[9]  
HOELEN CGA, 1998, OSA TOPS, V21, P386
[10]  
HOELEN CGA, 1997, P SOC PHOTO-OPT INS, V3196, P142