Immediate and Long-Term Echocardiographic Findings after Transcatheter Aortic Valve Implantation for the Treatment of Aortic Stenosis: The Cribier-Edwards/Edwards-Sapien Valve Experience

被引:36
作者
Bauer, Fabrice [1 ,2 ]
Lemercier, Mathieu [1 ]
Zajarias, Alan
Tron, Christophe
Eltchaninoff, Helene
Cribier, Alain [1 ,2 ]
机构
[1] Rouen Univ Hosp, Div Cardiol, Rouen, France
[2] INSERM, U644, Rouen Med Sch, Res Lab, Rouen, France
关键词
Aortic stenosis; Doppler; Prosthesis; Surgery; Transcatheter heart valve implantation; PERCUTANEOUS BALLOON VALVULOPLASTY; PROSTHETIC HEART-VALVE; FOLLOW-UP; REPLACEMENT; REGURGITATION; IMPROVEMENT; IMPACT;
D O I
10.1016/j.echo.2010.01.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The role of transcatheter aortic valve implantation in the treatment of calcific aortic stenosis is evolving. Immediate and long-term echocardiographic findings are poorly reported. Methods: Eighty-eight patients in whom surgical aortic valve replacement was contraindicated were studied before and 1 and 7 days, 1 month, and 1 and 2 years after the transcatheter procedure by echocardiography for hemodynamic. Transaortic pressure gradient, permeability index, and aortic valve area were measured, and aortic regurgitation was estimated from a multiparametric approach. A subset group of 36 patients (23-mm valve, n = 18; 26-mm valve, n = 18) with optimal ultrasound window were investigated for valve geometry at 7 days. We measured the sphericity index (anteroposterior to sagittal diameter ratio) and the angulation of the prosthesis with the ascending aorta. Results: By analysis of variance, transaortic pressure gradient significantly decreased and aortic valve area increased after the procedure (P < .0001 and P < .0001 respectively). Aortic regurgitation severity tended to decline at follow-up (P = .20) and was unaffected by valve size (P = .35). Leaks were paraprosthetic in 77% of cases, intraprosthetic in 6% of cases, and both in 17% of cases. Overall, the sphericity index was 1.03 +/- 0.07 and the angulation was 2.9 +/- 1.1 degrees. Conclusion: Echocardiography aids in the demonstration of appropriate prosthesis function and positioning after transcatheter aortic valve implantation. (J Am Soc Echocardiogr 2010;23:370-6.)
引用
收藏
页码:370 / 376
页数:7
相关论文
共 31 条
  • [1] Acute improvement in global and regional left ventricular systolic function after percutaneous heart valve implantation in patients with symptomatic aortic stenosis
    Bauer, F
    Eltchaninoff, H
    Tron, C
    Lesault, PF
    Agatiello, C
    Nercolini, D
    Derumeaux, G
    Cribier, A
    [J]. CIRCULATION, 2004, 110 (11) : 1473 - 1476
  • [2] PERCUTANEOUS BALLOON VALVULOPLASTY IN PATIENTS WITH SEVERE AORTIC-STENOSIS AND LOW EJECTION FRACTION - IMMEDIATE RESULTS AND 1-YEAR FOLLOW-UP
    BERLAND, J
    CRIBIER, A
    SAVIN, T
    LEFEBVRE, E
    KONING, R
    LETAC, B
    [J]. CIRCULATION, 1989, 79 (06) : 1189 - 1196
  • [3] DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION
    CHAFIZADEH, ER
    ZOGHBI, WA
    [J]. CIRCULATION, 1991, 83 (01) : 213 - 223
  • [4] Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis
    Cribier, A
    Eltchaninoff, H
    Tron, C
    Bauer, F
    Agatiello, C
    Sebagh, L
    Bash, A
    Nusimovici, D
    Litzler, PY
    Bessou, JP
    Leon, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (04) : 698 - 703
  • [5] Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description
    Cribier, A
    Eltchaninoff, H
    Bash, A
    Borenstein, N
    Tron, C
    Bauer, F
    Derumeaux, G
    Anselme, F
    Laborde, F
    Leon, MB
    [J]. CIRCULATION, 2002, 106 (24) : 3006 - 3008
  • [6] CRIBIER A, 1988, HERZ, V13, P110
  • [7] Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis
    Dellgren, G
    David, TE
    Raanani, E
    Armstrong, S
    Ivanov, J
    Rakowski, H
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (01) : 146 - 154
  • [8] Determinants of Significant Paravalvular Regurgitation After Transcatheter Aortic Valve Implantation Impact of Device and Annulus Discongruence
    Detaint, Delphine
    Lepage, Laurent
    Himbert, Dominique
    Brochet, Eric
    Messika-Zeitoun, David
    Iung, Bernard
    Vahanian, Alec
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (09) : 821 - 827
  • [9] Eltchaninoff Helene, 2003, J Interv Cardiol, V16, P515, DOI 10.1046/j.1540-8183.2003.01051.x
  • [10] FRIEDMAN SA, 1970, NEW ENGL J MED, V282, P396