Thermal dose optimization via temporal switching in ultrasound surgery

被引:70
作者
Daum, DR
Hynynen, K
机构
[1] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Boston, MA 02115 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/58.646926
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Temporal switching has been simulated and implemented in vivo experiments as a method to optimize thermal dose in ultrasound surgery. By optimizing the thermal dose over a tissue volume, the peak temperature is decreased, less average power is expended, and overall treatment time is shortened. To test this hypothesis, a 16 element, spherically sectioned array has been constructed for application in ultrasound surgery guided by magnetic resonance imaging. A simulation study for the array was performed to determine an optimal treatment from a set of multiple focus fields. These fields were generated using the mode scanning technique with power levels determined numerically using a direct weighted gradient search in the attempt to create an optimally uniform thermal dose over a 0.6 X 0.6 X 1.0 cm(3) tissue volume. Comparisons of the switched fields and a static multiple focus field indicate that the switching technique can lower power requirements and decrease treatment time by 20%. More importantly, the peak temperature of the sonication was lowered 13 degrees C, thus decreasing the possibility of cavitation. The simulated results of the 16 element array were then experimentally tested using MRI to noninvasively monitor temperature elevations and predict lesion size in rabbit thigh muscle in vivo. In addition, the results show that the switching technique can be less sensitive to tissue inhomogeneities than static field sonication while creating contiguous necrosis regions at equal average powers.
引用
收藏
页码:208 / 215
页数:8
相关论文
共 25 条
[1]  
[Anonymous], 2012, ESSENTIALS MED ULTRA
[2]   CONCENTRIC-RING AND SECTOR-VORTEX PHASED-ARRAY APPLICATORS FOR ULTRASOUND HYPERTHERMIA [J].
CAIN, CA ;
UMEMURA, S .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :542-551
[3]   Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam [J].
Chung, AH ;
Hynynen, K ;
Colucci, V ;
Oshio, K ;
Cline, HE ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :745-752
[4]   FOCAL SPACING AND NEAR-FIELD HEATING DURING PULSED HIGH-TEMPERATURE ULTRASOUND THERAPY [J].
DAMIANOU, C ;
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :777-787
[5]   THE EFFECTS OF TISSUE HETEROGENEITIES AND LARGE BLOOD-VESSELS ON THE THERMAL EXPOSURE INDUCED BY SHORT HIGH-POWER ULTRASOUND PULSES [J].
DORR, LN ;
HYNYNEN, K .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (01) :45-59
[7]   A SPHERICAL-SECTION ULTRASOUND PHASED-ARRAY APPLICATOR FOR DEEP LOCALIZED HYPERTHERMIA [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (07) :634-643
[8]   OPTIMIZATION OF THE INTENSITY GAIN OF MULTIPLE-FOCUS PHASED-ARRAY HEATING PATTERNS [J].
EBBINI, ES ;
CAIN, CA .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1991, 7 (06) :953-973
[9]   Ultrasound surgery using multiple sonications - Treatment time considerations [J].
Fan, XB ;
Hynynen, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (04) :471-482
[10]   A study of various parameters of spherically curved phased arrays for noninvasive ultrasound surgery [J].
Fan, XB ;
Hynynen, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (04) :591-608