Determination of cadmium by flow injection isotope dilution inductively coupled plasma mass spectrometry with vapour generation sample introduction

被引:81
作者
Hwang, TJ [1 ]
Jiang, SJ [1 ]
机构
[1] NATL SUN YAT SEN UNIV,DEPT CHEM,KAOHSIUNG 80424,TAIWAN
关键词
inductively coupled plasma mass spectrometry; isotope dilution; cadmium; flow injection; vapour generation; biological and environmental samples;
D O I
10.1039/a607469c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and inexpensive laboratory-built vapour generator was used with isotope dilution-ICP-MS for the determination of cadmium in several reference materials, The application of vapour generation-ICP-MS alleviated the spectroscopic interferences in cadmium determination encountered when a conventional pneumatic nebulizer was used for sample introduction, The sensitivity, detection limit and repeatability of the flow injection-ICP-MS system with a vapour generator were comparable to those of ICP-MS analysis with conventional pneumatic nebulization. The repeatability of the peak areas and isotope ratio determinations of seven consecutive injections of a 1 ng ml(-1) cadmium solution was 3.5 and 1.9%, respectively, The method has a detection limit of 0.026 ng ml(-1) for cadmium, The method was applied to the determination of cadmium in NIST SRM 2704 and National Research Council of Canada (NRCC) BCCS-1 and PACS-1 sediment reference materials, NIST SRM 2670 Toxic Metals in Freeze-Dried Urine and several water reference materials, The concentrations of cadmium in the reference materials were quantified by the isotope dilution method. The precision between sample replicates was better than 10% and the analysis results were within 5% of the certified values for most of the determinations.
引用
收藏
页码:579 / 584
页数:6
相关论文
共 37 条
[1]   A COMPARISON OF ISOTOPE-DILUTION MASS-SPECTROMETRIC METHODS FOR THE ASSAY OF COPPER IN COPPER ORE REFERENCE MATERIALS [J].
BEARY, ES ;
BRLETIC, KA ;
PAULSEN, PJ ;
MOODY, JR .
ANALYST, 1987, 112 (04) :441-444
[2]   DETERMINATION OF TRACE-METALS IN AN OPEN OCEAN WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (02) :305-308
[3]   DETERMINATION OF URANIUM IN ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BOOMER, DW ;
POWELL, MJ .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2810-2813
[4]   ELECTROTHERMAL ISOTOPE-DILUTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY METHOD FOR THE DETERMINATION OF SUB-NG ML(-1) LEVELS OF LEAD IN HUMAN PLASMA [J].
BOWINS, RJ ;
MCNUTT, RH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (11) :1233-1236
[5]   COMBINED GENERATOR/SEPARATOR FOR CONTINUOUS HYDRIDE GENERATION - APPLICATION TO ONLINE PRE-REDUCTION OF ARSENIC(V) AND DETERMINATION OF ARSENIC IN WATER BY ATOMIC EMISSION-SPECTROMETRY [J].
BRINDLE, ID ;
ALARABI, H ;
KARSHMAN, S ;
LE, XC ;
ZHENG, SG ;
CHEN, HW .
ANALYST, 1992, 117 (03) :407-411
[6]   GENERATION OF A VOLATILE CADMIUM SPECIES IN AN ORGANIC MEDIUM [J].
CACHO, J ;
BELTRAN, I ;
NERIN, C .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (07) :661-663
[7]   COMBINED GENERATOR SEPARATOR .2. STIBINE GENERATION COMBINED WITH FLOW-INJECTION FOR THE DETERMINATION OF ANTIMONY IN METAL SAMPLES BY ATOMIC EMISSION-SPECTROMETRY [J].
CHEN, HW ;
BRINDLE, ID ;
ZHENG, SG .
ANALYST, 1992, 117 (10) :1603-1608
[8]   SELECTION OF MODE FOR THE MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY AND ITS APPLICATION TO MILK POWDER ANALYSIS [J].
DEAN, JR ;
EBDON, L ;
MASSEY, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (04) :369-374
[9]   DETERMINATION OF CADMIUM IN AQUEOUS SAMPLES BY VAPOR GENERATION WITH SODIUM TETRAETHYLBORATE(III) REAGENT [J].
DULIVO, A ;
CHEN, YW .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (04) :319-322
[10]   ONLINE REMOVAL OF INTERFERENCES IN THE ANALYSIS OF BIOLOGICAL-MATERIALS BY FLOW-INJECTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
EBDON, L ;
FISHER, AS ;
WORSFOLD, PJ ;
CREWS, H ;
BAXTER, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (05) :691-695