Investigation of tomotherapy beam delivery

被引:106
作者
Yang, JN [1 ]
Mackie, TR [1 ]
Reckwerdt, P [1 ]
Deasy, JO [1 ]
Thomadsen, BR [1 ]
机构
[1] UNIV WISCONSIN,DEPT HUMAN ONCOL,MADISON,WI 53706
关键词
tomotherapy; conformal radiotherapy; helical beam; motion artifacts;
D O I
10.1118/1.597909
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Experimental simulations for tomotherapy beam delivery were performed using a computer-controlled phantom positioner, a cylindrical phantom, and a 6 MV x-ray slit beam. Both continuous helical beam and sequential segmented tomotherapy (SST) beam deliveries were evaluated. Beam junctioning problem due to couch indexing error or field width errors presented severe dose uniformity perturbations for SST, while the problem was minimized for helical beam delivery. Longitudinal breathing motions were experimentally simulated for helical and SST beam delivery. While motions reduced the dose uniformity perturbations for SST, small artifacts in dose uniformity can be introduced for helical beam delivery. With typical breath frequency and magnitude, for a slit beam of 2.0 cm width at 4 rpm, the dose uniformity perturbation was not significant. A running start/stop technique was implemented with helical beam delivery to sharpen the 20%-80% longitudinal dose fall-off from 1.5 to 0.5 cm. The latter was comparable to the corresponding dose penumbra of a conventional 6 MV 10x10 cm(2) field. All together, helical beam delivery showed advantages over SST for tomotherapy beam delivery under similar delivery conditions. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:425 / 436
页数:12