How does the inclination of the tibial component matter? A three-dimensional finite element analysis of medial mobile-bearing unicompartmental arthroplasty

被引:48
作者
Dai, Xuesong [1 ]
Fang, Jinghua [1 ]
Jiang, Lifeng [1 ]
Xiong, Yan [1 ]
Zhang, Miaofeng [1 ]
Zhu, Sunan [1 ]
机构
[1] Zhejiang Univ, Affiliated Hosp 2, Dept Orthopaed, Sch Med, Hangzhou, Zhejiang, Peoples R China
关键词
UKA; Tibial stress; Varus; Valgus; Finite element analysis; UNICONDYLAR KNEE ARTHROPLASTY; EXPERIMENTAL VALIDATION; SURVIVORSHIP; REPLACEMENT; OUTCOMES; MODEL;
D O I
10.1016/j.knee.2018.02.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Medial unicompartmental knee arthroplasty (UKA) using Oxford mobile-bearing prosthesis is performed in the treatment of medial compartmental arthritis of the knee. However, little is known about the stress distributions for mobile-bearing UKA on the medial tibial plateau. Methods: In this study, the stresses on the coronal plane were calculated in a three-dimensional model of the proximal tibia. The features of the stress distribution were investigated when the tibial tray was placed in 15 degrees, 10 degrees, six degrees, and three degrees varus, neutral (0 degrees), and in three degrees, six degrees, 10 degrees, and 15 degrees valgus on the coronal plane of the medial plateau. Results: The peak von Mises stress was found on the cortex below the medial plateau while the stresses of cortical bone increased gradually as the inclination of the tibial tray was changed from varus to valgus. The amount of peak stress was almost the same as that in the normal knee model when the tibial tray was placed in six degrees valgus and consistently lower in varus inclination than in the normal knee model. Conversely, the peak stress of soft bone was found at the bottom of the slot. Conclusions: This study demonstrates that the inclination of the tibial component affects stress distribution in the proximal tibia after UKA. Slight varus inclination of the mobile-bearing tibial component is acceptable as it lowers the peak stress on the medial cortex. Additionally, placing the tibial tray in slight varus avoids a rise in stress between the tip of the keel and the medial tibial cortex. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:434 / 444
页数:11
相关论文
共 26 条
[1]
Survivorship and patient satisfaction of a fixed bearing unicompartmental knee arthroplasty incorporating an all-polyethylene tibial component [J].
Bhattacharya, R. ;
Scott, C. E. H. ;
Morris, H. E. ;
Wade, F. ;
Nutton, R. W. .
KNEE, 2012, 19 (04) :348-351
[2]
Muscle activity around the knee and gait performance in unicompartmental knee arthroplasty patients: a comparative study on fixed- and mobile-bearing designs [J].
Catani, Fabio ;
Benedetti, Maria Grazia ;
Bianchi, Luca ;
Marchionni, Valentina ;
Giannini, Sandro ;
Leardini, Alberto .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (06) :1042-1048
[3]
Biomechanical evaluation of proximal tibial behavior following unicondylar knee arthroplasty: Modified resected surface with corresponding surgical technique [J].
Chang, Tsung-Wei ;
Yang, Chan-Tsung ;
Liu, Yu-Liang ;
Chen, Wen-Chuan ;
Lin, Kun-Jhih ;
Lai, Yu-Shu ;
Huang, Chang-Hung ;
Lu, Yung-Chang ;
Cheng, Cheng-Kung .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (10) :1175-1182
[4]
Fixed-bearing unicompartmental knee arthroplasty. Patients' selection and operative technique [J].
Deschamps, G. ;
Chol, C. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (06) :648-661
[5]
Coronal Alignment in Total Knee Arthroplasty Just How Important is it? [J].
Fang, David M. ;
Ritter, Merrill A. ;
Davis, Kenneth E. .
JOURNAL OF ARTHROPLASTY, 2009, 24 (06) :39-43
[6]
Fixed or mobile bearing unicompartmental knee replacement? A comparative cohort study [J].
Gleeson, RE ;
Evans, R ;
Ackroyd, CE ;
Webb, J ;
Newman, JH .
KNEE, 2004, 11 (05) :379-384
[7]
Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis [J].
Godest, AC ;
Beaugonin, M ;
Haug, E ;
Taylor, M ;
Gregson, PJ .
JOURNAL OF BIOMECHANICS, 2002, 35 (02) :267-275
[8]
Experimental validation of a finite element model of a composite tibia [J].
Gray, H. A. ;
Zavatsky, A. B. ;
Taddei, F. ;
Cristofolini, L. ;
Gill, H. S. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H3) :315-324
[9]
Experimental validation of a finite element model of a human cadaveric tibia [J].
Gray, Hans A. ;
Taddei, Fulvia ;
Zavatsky, Amy B. ;
Cristofolini, Luca ;
Gill, Harinderjit S. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2008, 130 (03)
[10]
Survivorship of UKA in the middle-aged [J].
Heyse, Thomas J. ;
Khefacha, Ahmed ;
Peersman, Geert ;
Cartier, Philippe .
KNEE, 2012, 19 (05) :585-591