Evaluation and comparison of very high pressure liquid chromatography systems for the separation and validation of pharmaceutical compounds

被引:30
作者
Cunliffe, Jennifer M. [1 ]
Adams-Hall, Sharee B. [1 ]
Maloney, Todd D. [1 ]
机构
[1] Pfizer Global Res & Dev, Michigan Labs, Ann Arbor, MI 48105 USA
关键词
HPLC; sub-2 mu m; very high pressure liquid chromatography;
D O I
10.1002/jssc.200600524
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Chromatography using sub-2 mu m. particles is becoming increasingly popular due to the potential for increased speed, resolution, sensitivity, and peak capacity. To meet the demand, various vendors have re-engineered traditional LC systems to operate at pressures of up to 15 000 psi to accommodate the elevated backpressures associated with using sub-2 mu m particles. This report investigates and compares the performance of three very high pressure LC (VHPLC) systems: Waters Acquity, Agilent 1200 SL, and Thermo Accela. Specifications for the pump, autosampler, column compartment, detector, and software for each instrument are presented. To assess the chromatographic performance of the three instruments, method development and validation were performed for three pharmaceutical compounds and the results are compared and discussed. The material presented herein serves to highlight the different features of the VHPLC instruments, and assess their suitability for the analysis of pharmaceutical compounds.
引用
收藏
页码:1214 / 1223
页数:10
相关论文
共 23 条
[1]   Injection valve for ultrahigh-pressure liquid chromatography [J].
Anspach, JA ;
Maloney, TD ;
Brice, RW ;
Colón, LA .
ANALYTICAL CHEMISTRY, 2005, 77 (22) :7489-7494
[2]   Product ion mass spectra of amphetamine-type substances, designer analogues, and ketamine using ultra-performance liquid chromatography/tandem mass spectrometry [J].
Apollonio, Luigino G. ;
Whittall, Ian R. ;
Pianca, Dennis J. ;
Kyd, Jennelle M. ;
Maher, William A. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (15) :2259-2264
[3]   Very high pressure HPLC with 1 mm id columns [J].
Colón, LA ;
Cintrón, JM ;
Anspach, JA ;
Fermier, AM ;
Swinney, KA .
ANALYST, 2004, 129 (06) :503-504
[4]  
Jerkovich A.D., 2003, LC GC N AM, V21, P600
[5]   Investigating the human metabolism of acetaminophen using UPLC and exact mass oa-TOF MS [J].
Johnson, KA ;
Plumb, R .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 39 (3-4) :805-810
[6]   Rapid determination of histamine and its metabolites in mice hair by ultra-performance liquid chromatography with time-of-flight mass spectrometry [J].
Kawanishi, Hiroki ;
Toyo'oka, Toshimasa ;
Ito, Kenichi ;
Maeda, Munenori ;
Hamada, Tomoj I. ;
Fukushima, Takeshi ;
Kato, Masaru ;
Inagaki, Shinsuke .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1132 (1-2) :148-156
[7]   Comparison of ultra-performance liquid chromatography and high-performance liquid chromatography for the determination of priority pesticides in baby foods by tandem quadrupole mass spectrometry [J].
Leandro, CC ;
Hancock, P ;
Fussell, RJ ;
Keely, BJ .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1103 (01) :94-101
[8]  
Lippert JA, 1999, J MICROCOLUMN SEP, V11, P631, DOI 10.1002/(SICI)1520-667X(199911)11:9<631::AID-MCS1>3.3.CO
[9]  
2-9
[10]   Ultrahigh pressure reversed-phase capillary liquid chromatography:: Isocratic and gradient elution using columns packed with 1.0-μm particles [J].
MacNair, JE ;
Patel, KD ;
Jorgenson, JW .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :700-708