Determination of arsenic(III) and arsenic(V) by electrothermal atomic absorption spectrometry after-complexation and sorption on a C-18 bonded silica column

被引:84
作者
Pozebon, D [1 ]
Dressler, VL [1 ]
Neto, JAG [1 ]
Curtius, AJ [1 ]
机构
[1] Univ Fed Santa Catarina, Dept Quim, BR-88040900 Florianopolis, SC, Brazil
关键词
arsenic; C-18 bonded silica; diethyl dithiophosphate; flow injection analysis; electrothermal atomic absorption spectrometry;
D O I
10.1016/S0039-9140(97)00234-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection procedure for the separation and pre-concentration of inorganic arsenic based on the complexation with ammonium diethyl dithiophosphate (DDTP) and sorption on a C-18 bonded silica gel minicolumn is proposed. During the sample injection by a time-based fashion, the As3+-DDTP complex is stripped from the solution and retained in the column. Arsenic(V) and other ions that do not form complexes are discarded. After reduction to the trivalent slate by using potassium iodide plus ascorbic acid, total arsenic is determined by electrothermal atomic absorption spectrometry (ETAAS). Arsenic(V) concentration can be calculated by difference. After processing 6 ml sample volume, the As3+-DDTP complexes were eluted directly into the autosampler cup (120 mu l). Ethanol was used for column rinsing. Influence of pH, reagent concentration, pre-concentration and elution time and column size were investigated. When 30 mu l of eluate plus 10 mu l of 0.1% (w/v) Pd(NO3)(2) were dispensed into the graphite tube, analytical curve in the 0.3-3 mu g As 1(-1) range was obtained (r = 0.9991). The accuracy was checked for arsenic determination in a certified water, spiked tap water and synthetic mixtures of arsenite and arsenate. Good recoveries (97-108%) of spiked samples were found. Results are precise (RSD 7.5 and 6%, for 0.5 and 2.5 mu g 1(-1), n = 10) and in agreement with the certified value of reference material at 95% confidence level. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:1167 / 1175
页数:9
相关论文
共 15 条
[1]  
Batley GE, 1989, TRACE ELEMENTS SPECI
[2]   SPECIATION OF ARSENIC BY THE DETERMINATION OF TOTAL ARSENIC AND ARSENIC(III) IN MARINE SEDIMENT SAMPLES BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BERMEJOBARRERA, P ;
BARCIELAALONSO, MC ;
FERRONNOVAIS, M ;
BERMEJOBARRERA, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (03) :247-252
[3]  
BODE H, 1962, Z ANAL CHEM, V185, P179
[4]   THE DETERMINATION OF ARSENIC BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY WITH A GRAPHITE-FURNACE .2. DETERMINATION OF ARSENIC(III) AND ARSENIC(V) AFTER EXTRACTION [J].
CHAKRABORTI, D ;
DEJONGHE, W ;
ADAMS, F .
ANALYTICA CHIMICA ACTA, 1980, 120 (NOV) :121-127
[5]   URINARY ARSENIC SPECIATION BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ATOMIC-ABSORPTION SPECTROMETRY FOR MONITORING OCCUPATIONAL EXPOSURE TO INORGANIC ARSENIC [J].
CHANA, BS ;
SMITH, NJ .
ANALYTICA CHIMICA ACTA, 1987, 197 :177-186
[6]   DIRECT COUPLING OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY TO MICROWAVE-INDUCED PLASMA-ATOMIC EMISSION-SPECTROMETRY VIA VOLATILE-SPECIES GENERATION AND ITS APPLICATION TO MERCURY AND ARSENIC SPECIATION [J].
COSTAFERNANDEZ, JM ;
LUNZER, F ;
PEREIROGARCIA, R ;
SANZMEDEL, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (11) :1019-1025
[7]  
IRGOLIC JK, 1985, ENV INORGANIC CHEM
[8]  
MA R, 1996, SPECTROCHIM ACTA B, V51
[9]   Determination of ultra-trace amounts of arsenic(III) by flow-injection hydride generation atomic absorption spectrometry with on-line preconcentration by coprecipitation with lanthanum hydroxide or hafnium hydroxide [J].
Nielsen, S ;
Sloth, JJ ;
Hansen, EH .
TALANTA, 1996, 43 (06) :867-880
[10]  
SPERLING M, 1991, SPECTROCHIM ACTA B, V46, P1798