A new miniature x-ray device for interstitial radiosurgery: Dosimetry

被引:99
作者
Beatty, J [1 ]
Biggs, PJ [1 ]
Gall, K [1 ]
Okunieff, P [1 ]
Pardo, FS [1 ]
Harte, KJ [1 ]
Dalterio, MJ [1 ]
Sliski, AP [1 ]
机构
[1] PHOTOELECTRON CORP,WALTHAM,MA 02154
关键词
stereotactic radiosurgery; dosimetry; low-energy x rays;
D O I
10.1118/1.597791
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A miniature, battery operated 40 kV x-ray device has been developed for the interstitial treatment of small tumors (<3 cm diam) in humans. X rays are emitted from the tip of a 10 cm long, 3 mm diameter probe that is stereotactically inserted into the tumor. The beam, characterized by half-value layer (HVL), spectrum analysis, and isodose contours, behaves essentially as a point isotropic source with an effective energy of 20 keV at a depth of 10 mm in water. The absolute output from the device was measured using a parallel plate ionization chamber, modified with a platinum aperture. The dose rate in water determined from these chamber measurements was found to be nominally 150 cGy/min at a distance of 10 mm for a beam current of 40 mu A and voltage of 40 kV. The dose in water falls off approximately as the third power of the distance. To date, 14 patients have been treated with this device in a phase I clinical trial. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:53 / 62
页数:10
相关论文
共 9 条
[1]  
APUZZO MLJ, 1988, STEREOTACTIC NEUROSU, V2
[2]  
COSGROVE GR, IN PRESS TXB STEREOT
[3]  
FLICKINGER JC, 1992, STEREOTACTIC RADIOSU, V3, P51
[4]  
GALL K, 1993, MED PHYS, V20, P925
[5]  
JOHNS HE, 1974, PHYSICS RADIOLOGY, P303
[6]   THE INFLUENCE OF VOLUME ON THE TOLERANCE OF THE BRAIN TO RADIOSURGERY [J].
MARKS, LB ;
SPENCER, DP .
JOURNAL OF NEUROSURGERY, 1991, 75 (02) :177-180
[7]   A COMPARISON OF SOLID PHANTOMS WITH WATER FOR DOSIMETRY OF I125 BRACHYTHERAPY SOURCES [J].
MEIGOONI, AS ;
MELI, JA ;
NATH, R .
MEDICAL PHYSICS, 1988, 15 (05) :695-701
[8]  
SMITH D, 1993, MED PHYS, V20, P1291
[9]  
1970, ICRU17 REP