Reconstruction of axial tomographic high resolution data from confocal fluorescence microscopy: a method for improving 3D FISH images

被引:21
作者
Heintzmann, R
Kreth, G
Cremer, C
机构
[1] Univ Heidelberg, Inst Appl Phys, D-69120 Heidelberg, Germany
[2] Univ Heidelberg, Interdisciplinary Ctr Sci Comp, Heidelberg, Germany
来源
ANALYTICAL CELLULAR PATHOLOGY | 2000年 / 20卷 / 01期
关键词
D O I
10.1155/2000/459351
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Fluorescent confocal laser scanning microscopy allows an improved imaging of microscopic objects in three dimensions. However, the resolution along the axial direction is three times worse than the resolution in lateral directions. A method to overcome this axial limitation is tilting the object under the microscope, in a way that the direction of the optical axis points into different directions relative to the sample. A new technique for a simultaneous reconstruction from a number of such axial tomographic confocal data sets was developed and used for high resolution reconstruction of 3D-data both from experimental and virtual microscopic data sets. The reconstructed images have a highly improved 3D resolution, which is comparable to the lateral resolution of a single deconvolved data set. Axial tomographic imaging in combination with simultaneous data reconstruction also opens the possibility for a more precise quantification of 3D data. The color images of this publication can be accessed from http://www.esacp.org/acp/2000/20-1/heintzmann.htm. At this web address an interactive 3D viewer is additionally provided for browsing the 3D data. This java applet displays three orthogonal slices of the data set which are dynamically updated by user mouse clicks or keystrokes.
引用
收藏
页码:7 / 15
页数:9
相关论文
共 18 条
[1]  
Bradl J, 1996, P SOC PHOTO-OPT INS, V2628, P140
[2]  
BRADL J, 1996, MICROSCOPY ANAL NOV, P9
[3]  
Canterakis Nikolaos, 1996, MUST 18 DAGM S INF A, P339
[4]   ELLIPTIC FIT OF OBJECTS IN 2 AND 3 DIMENSIONS BY MOMENT OF INERTIA OPTIMIZATION [J].
CHAUDHURI, BB ;
SAMANTA, GP .
PATTERN RECOGNITION LETTERS, 1991, 12 (01) :1-7
[5]   Fluorescence microtomography: Multi-angle image acquisition and 3D digital reconstruction [J].
Cogswell, CJ ;
Larkin, KG ;
Klemm, HU .
THREE-DIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING III, 1996, 2655 :109-115
[6]   ABERRATIONS IN CONFOCAL FLUORESCENCE MICROSCOPY INDUCED BY MISMATCHES IN REFRACTIVE-INDEX [J].
HELL, S ;
REINER, G ;
CREMER, C ;
STELZER, EHK .
JOURNAL OF MICROSCOPY, 1993, 169 :391-405
[7]  
HOLMES TJ, 1995, HDB BIOL CONFOCAL MI, pCH24
[8]  
Jahne Bernd, 1999, HDB COMPUTER VISION, V1
[9]  
KAWATA S, 1994, ADV OPTICAL ELECT MI, V14, P213
[10]  
KRETH G, 2000, IN PRESS CHROMOSOME