Efficacy and biocompatibility of a photopolymerized, synthetic, absorbable hydrogel as a dural sealant in a canine craniotomy model

被引:35
作者
Alleyne, CH
Cawley, CM
Barrow, DL
Poff, BC
Powell, MD
Sawhney, AS
Dillehay, DL
机构
[1] Emory Univ, Sch Med, Dept Neurosurg, Atlanta, GA 30322 USA
[2] Emory Univ, Sch Med, Dept Pathol, Atlanta, GA 30322 USA
[3] Emory Univ, Sch Med, Div Anim Resources, Atlanta, GA 30322 USA
[4] Focal Inc, Lexington, MA USA
关键词
cerebrospinal fluid leak; dural defect repair; dural sealant; dural tear; hydrogel; meninges; dog;
D O I
10.3171/jns.1998.88.2.0308
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. A canine craniotomy model was used to evaluate the dural sealing efficacy and biocompatibility of a novel, synthetic, bioresorbable hydrogel. Methods. Bilateral craniotomies were performed in 24 dogs assigned to six survival periods. In each animal a parasagittal durotomy was created and then repaired. At the treatment sites the hydrogel sealant was applied over the dural repair and photopolymerized. The repair was tested for leaks to 20 cm H2O by using a Valsalva maneuver. At the control sites the incisions were sutured and tested for leaks only. After uneventful survival periods, the leak test was repeated in three of the four animals in each group. Bone-dura adhesion was evaluated, after which the dura and underlying brain were removed, fixed, and examined histologically. En bloc histological investigation was performed on a specimen obtained from the fourth animal in each group. Over a 56-day period, 18 treated sites were tested for leaks. A leak was detected at a site remote from that of the repair in one animal; this was excluded from analysis. Thus 17 of 17 treated sites remained free of leaks. On the control side of one animal. there was a leak from a new dural tear at the cranial end of the durotomy, which occurred when the bone flap was removed. This site was also excluded from analysis. Eleven of 17 leak-tested control sites remained free of leaks over the study period. Bone-dura adhesions occurred in 15 of 19 control sites and had a mean adhesion score of 1.37 (range 0-4), whereas adhesions occurred in 10 of 19 treated sites with a mean adhesion score of 0.84 (range 0-3). No cortical reaction was noted. Conclusions. This novel hydrogel sealant is efficacious in sealing dural repair sites measuring up to 2 mm. Healing of the underlying dura is not compromised and exposed cortical tissue is not altered histologically.
引用
收藏
页码:308 / 313
页数:6
相关论文
共 45 条
  • [1] Abbe R, 1895, T AM SURG ASS, V13, P490
  • [2] CLINICAL RESULTS OF LYOPHILIZED HUMAN CADAVER DURA TRANSPLANTATION
    ABBOTT, WM
    DUPREE, EL
    [J]. JOURNAL OF NEUROSURGERY, 1971, 34 (06) : 770 - &
  • [3] THE ROLE OF NEOMEMBRANES IN FORMATION OF HEMATOMA AROUND SILASTIC DURA SUBSTITUTE - CASE-REPORT
    ADEGBITE, AB
    PAINE, KWE
    ROZDILSKY, B
    [J]. JOURNAL OF NEUROSURGERY, 1983, 58 (02) : 295 - 297
  • [4] IMMUNE-RESPONSE IN HOSTS WITH CADAVERIC DURAL GRAFTS - REPORT OF 2 CASES
    ALLEYNE, CH
    BARROW, DL
    [J]. JOURNAL OF NEUROSURGERY, 1994, 81 (04) : 610 - 613
  • [5] Alleyne CH, 1996, NEUROSURGERY, V39, P768
  • [6] Bovine pericardium for dural crafts: Clinical results in 35 patients
    Alleyne, CH
    AlMefty, O
    [J]. NEUROSURGERY, 1996, 39 (04) : 768 - 768
  • [7] BANGZONG X, 1988, J NEUROSURG, V69, P707
  • [8] BEACH HHA, 1890, BOSTON MED SURG J, V122, P313
  • [9] A SYNTHETIC DURAL PROSTHESIS CONSTRUCTED FROM HYDROXYETHYLMETHACRYLATE HYDROGELS
    BHATIA, S
    BERGETHON, PR
    BLEASE, S
    KEMPER, T
    ROSIELLO, A
    ZIMBARDI, GP
    FRANZBLAU, C
    SPATZ, EL
    [J]. JOURNAL OF NEUROSURGERY, 1995, 83 (05) : 897 - 902
  • [10] SILASTIC DURAPLASTY IN PEDIATRIC-PATIENTS
    BOOP, FA
    CHADDUCK, WM
    [J]. NEUROSURGERY, 1991, 29 (05) : 785 - 788