CT-based Monte Carlo simulation tool for dosimetry planning and analysis

被引:172
作者
DeMarco, JJ [1 ]
Solberg, TD [1 ]
Smathers, JB [1 ]
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Jonsson Comprehens Canc Ctr, Los Angeles, CA 90024 USA
关键词
Monte Carlo; treatment planning; computed tomography; dosimetry; inhomogeneity;
D O I
10.1118/1.598167
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Los Alamos code MCNP4A (Monte Carlo N-Particle version 4A) is currently used to simulate a variety of problems ranging from nuclear reactor analysis to boron neutron capture therapy. A graphical user interface has been developed that automatically sets up the MCNP4A geometry and radiation source requirements for a three-dimensional Monte Carlo simulation using computed tomography data. The major drawback for this dosimetry system is the amount of time to obtain a statistically significant answer. A specialized patch file has been developed that optimizes photon particle transport and dose scoring within the standard MCNP4A lattice geometry. The transport modifications produce a performance increase (number of histories per minute) of approximately 4.7 based upon a 6 MV point source centered within a 30 x 30 x 30 cm(3) lattice water phantom and 1 x 1 x 1 mm(3) voxels. The dose scoring modifications produce a performance increase of approximately 470 based upon a tally section of greater than 1 x 10(4) lattice elements and a voxel size of 5 mm(3). Homogeneous and heterogeneous benchmark calculations produce good agreement with measurements using a standard water phantom and a high-and low-density heterogeneity phantom. The dose distribution from a typical mediastinum treatment planning setup is presented for qualitative analysis and comparison versus a conventional treatment planning system. (C) 1998 American Association of Physicists in Medicine.
引用
收藏
页码:1 / 11
页数:11
相关论文
共 42 条
[11]  
Hartmann Siantar C. L., 1995, Proceedings of the International Conference, Mathematics and Computations, Reactor Physics, and Environmental Analyses, P857
[12]  
HARTMANNSIANTAR CL, 1995, MED PHYS, V22, P994
[13]  
HARTMANNSIANTAR CL, 1996, MED PHYS, V23, P1128
[14]  
HENDRICKS JS, 1992, LAUR913456 LOS AL NA
[15]  
ICRP, 1975, ICRP PUBL, V23
[16]   X-RAY SOURCES OF MEDICAL LINEAR ACCELERATORS - FOCAL AND EXTRA-FOCAL RADIATION [J].
JAFFRAY, DA ;
BATTISTA, JJ ;
FENSTER, A ;
MUNRO, P .
MEDICAL PHYSICS, 1993, 20 (05) :1417-1427
[17]  
Landau L, 1944, J PHYS-USSR, V8, P201
[18]  
LAUGHLIN JS, 1991, INT J RADIAT ONCOL, V21, P3
[19]   MEASUREMENT OF THE SOURCE SIZE OF A 6-MV AND 18-MV RADIOTHERAPY LINAC [J].
LOEWENTHAL, E ;
LOEWINGER, E ;
BARAVRAHAM, E ;
BARNEA, G .
MEDICAL PHYSICS, 1992, 19 (03) :687-690
[20]   MONTE-CARLO MODEL OF PHOTON BEAMS USED IN RADIATION-THERAPY [J].
LOVELOCK, DMJ ;
CHUI, CS ;
MOHAN, R .
MEDICAL PHYSICS, 1995, 22 (09) :1387-1394