Sequential injection separation system with stopped-flow radiometric detection for automated analysis of 99Tc in nuclear waste

被引:48
作者
Egorov, O
O'Hara, MJ
Ruzicka, J
Grate, JW
机构
[1] Pacific NW Lab, Environm Mol Sci Lab, Richland, WA 99352 USA
[2] Univ Washington, Dept Chem, Seattle, WA 98195 USA
关键词
D O I
10.1021/ac971121t
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated procedure for the determination of Tc-99 in aged nuclear waste has been developed. Using advanced sequential injection (SI) analysis instrumentation, Tc-99(VII) is separated from radioactive and stable interferences using a TEVA resin column that selectively retains pertechnetate ion from dilute nitric acid solutions. The separated Tc-99 is eluted with 6 M nitric acid and quantified on-line with a now-through liquid scintillation detector. A stopped-flow technique has been optimized that improves the analysis precision and detection limit compared to continuous-flow detection, reduces consumption of liquid scintillation cocktail, and increases sample throughput by separating the next sample while the present sample is being counted. The detection limit is 30 pCi, or 2 ng, of Tc-99, using a 15-min stopped-now period. The analysis time is 40 min for the first sample and is reduced to 20 min for each subsequent sample. Processed nuclear waste samples from the Hanford site were successfully analyzed by this new method.
引用
收藏
页码:977 / 984
页数:8
相关论文
共 31 条