Sensing lead insulation defect resulting in a damage of the ICD pulse generator case

被引:7
作者
Gummert, J
Krauss, B
Hutschenreiter, W
Hambrecht, R
Mohr, FW
机构
[1] Univ Leipzig, Ctr Heart, Dept Cardiol, D-7010 Leipzig, Germany
[2] Univ Leipzig, Ctr Heart, Dept Cardiac Surg, D-7010 Leipzig, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 02期
关键词
transvenous implantable cardioverter defibrillators; pulse generator case;
D O I
10.1111/j.1540-8159.1998.tb00079.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Eight months after ICD implantation with an electrically active case the patient presented with ICD system sensing failure. Upon reoperation, we found art insulation defect of the proximal sensing lead and the generator case showed are marks suggesting a short circuit between the sensing lead and case. The generator was replaced, the original sensing lead insulated, and a new sensing lead inserted. Bench testing of the explanted generator showed a damaged internal circuitry. Proximal lead insulation defects combined with electrically active cases may result in damage of the case. The potential damage of internal circuitry warrants generator replacement.
引用
收藏
页码:478 / 479
页数:2
相关论文
共 3 条
  • [1] Implantable transvenous cardioverter defibrillator leads: The dark side
    Lawton, JS
    Wood, MA
    Gilligan, DM
    Stambler, BS
    Damiano, RJ
    Ellenbogen, KA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (09): : 1273 - 1278
  • [2] Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators
    Lawton, JS
    Ellenbogen, KA
    Wood, MA
    Stambler, BS
    Herre, JM
    Nath, S
    Bernstein, RC
    DiMarco, JP
    Haines, DE
    Szentpetery, S
    Baker, LD
    Damiano, RJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (06) : 647 - 651
  • [3] SPURIOUS DISCHARGES DUE TO LATE INSULATION BREAK IN ENDOCARDIAL SENSING LEADS FOR CARDIOVERTER-DEFIBRILLATORS
    PETERS, RW
    FOSTER, AH
    SHOROFSKY, SR
    NOLAN, DA
    GOLD, MR
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (03): : 478 - 481