Evaluation of the silver iontophoretic catheter in an animal model

被引:16
作者
Hachem, RY
Wright, KC
Zermeno, A
Bodey, GP
Raad, II
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Infect Dis Infect Control & Employee Hlth, Houston, TX 77030 USA
[2] Univ Texas, MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
关键词
animal; catheter; silver;
D O I
10.1016/S0142-9612(03)00185-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Silver iontophoretic catheters (SIC) were shown to be highly efficacious in preventing catheter infections in vitro and in a rabbit model (J. Infect. Dis. 173 (1996) 495). Furthermore, we sought to determine the safety and durability of SIC prior to use in humans. A total of 30 New Zealand white rabbits (3-4kg) were randomly assigned to one of three groups whereby SIC or Arrow Guard (AG) catheters were tunneled and inserted in the jugular vein. All animals were followed for 2-12 weeks after catheter implantation. Blood was collected from each rabbit for assessment of toxicity and determination of silver levels. The electrical current generated by each SIC was measured once daily. At the end of the follow-up period, tissue samples were collected from the skin surrounding the catheter, the lungs, spleen, and liver. Microscopically, none of the tissues examined from any of the animals showed evidence of silver deposits, silver toxicity, thermal or electrical injury. The silver levels in the animals that received the SIC ranged from 0.1 to 2.23 mug/l with a mean of 0.62 (+/-0.44 SD). In conclusion SIC were safe with normal serum silver levels and were not associated with any local or systemic toxicity. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:3619 / 3622
页数:4
相关论文
共 10 条
[1]   ELECTRICALLY GENERATED SILVER IONS - QUANTITATIVE EFFECTS ON BACTERIAL AND MAMMALIAN-CELLS [J].
BERGER, TJ ;
SPADARO, JA ;
CHAPIN, SE ;
BECKER, RO .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1976, 9 (02) :357-358
[2]   MECHANISM OF ELECTRICAL ENHANCEMENT OF EFFICACY OF ANTIBIOTICS IN KILLING BIOFILM BACTERIA [J].
COSTERTON, JW ;
ELLIS, B ;
LAM, K ;
JOHNSON, F ;
KHOURY, AE .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1994, 38 (12) :2803-2809
[3]   A comparison of two antimicrobial-impregnated central venous catheters [J].
Darouiche, RO ;
Raad, II ;
Heard, SO ;
Thornby, JI ;
Wenker, OC ;
Gabrielli, A ;
Berg, J ;
Khardori, N ;
Hanna, H ;
Hachem, R ;
Harris, RL ;
Mayhall, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 340 (01) :1-8
[4]   THE EFFECTS OF ELECTRIC-CURRENT ON BACTERIA COLONIZING INTRAVENOUS CATHETERS [J].
LIU, WK ;
TEBBS, SE ;
BYRNE, PO ;
ELLIOTT, TSJ .
JOURNAL OF INFECTION, 1993, 27 (03) :261-269
[5]   Prevention of central venous catheter-related bloodstream infection by use of an antiseptic-impregnated catheter - A randomized, controlled trial [J].
Maki, DG ;
Stolz, SM ;
Wheeler, S ;
Mermel, LA .
ANNALS OF INTERNAL MEDICINE, 1997, 127 (04) :257-+
[6]   In vitro antimicrobial efficacy of silver iontophoretic catheter [J].
Raad, I ;
Hachem, R ;
Zermeno, A ;
Dumo, M ;
Bodey, GP .
BIOMATERIALS, 1996, 17 (11) :1055-1059
[7]   Silver iontophoretic catheter: A prototype of a long-term antiinfective vascular access device [J].
Raad, I ;
Hachem, R ;
Zermeno, A ;
Stephens, LC ;
Bodey, GP .
JOURNAL OF INFECTIOUS DISEASES, 1996, 173 (02) :495-498
[8]   Central venous catheters coated with minocycline and rifampin for the prevention of catheter-related colonization and bloodstream infections - A randomized, double-blind trial [J].
Raad, I ;
Darouiche, R ;
Dupuis, J ;
AbiSaid, D ;
Gabrielli, A ;
Hachem, R ;
Wall, M ;
Harris, R ;
Jones, J ;
Buzaid, A ;
Robertson, C ;
Shenaq, S ;
Curling, P ;
Burke, T ;
Ericsson, C ;
Greenberg, S ;
Hanania, N ;
Yosher, D ;
Gibson, D ;
Reardon, M ;
Reardon, P ;
Darnule, T ;
Mansouri, M ;
Rolston, K ;
Whimbey, E ;
Bivins, C ;
Huaringa, A ;
Price, K ;
Safar, H .
ANNALS OF INTERNAL MEDICINE, 1997, 127 (04) :267-+
[9]   Antimicrobial durability and rare ultrastructural colonization of indwelling central catheters coated with minocycline and rifampin [J].
Raad, II ;
Darouiche, RO ;
Hachem, R ;
Abi-Said, D ;
Safar, H ;
Darnule, T ;
Mansouri, M ;
Morck, D .
CRITICAL CARE MEDICINE, 1998, 26 (02) :219-224
[10]  
*THOMS HEALTHC SER, 1974, MICROMEDEX, V112