Determination of trace amount of lead in natural water by isotope dilution - Inductively coupled plasma mass spectrometry

被引:5
作者
Itoh, A [1 ]
Ozaki, T [1 ]
Chiba, K [1 ]
Haraguchi, H [1 ]
机构
[1] Nagoya Univ, Sch Engn, Dept Appl Chem, Chikusa Ku, Nagoya, Aichi 46401, Japan
关键词
ICP-MS; isotope dilution; lead; natural water;
D O I
10.2116/analsci.13.Supplement_7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Isotope dilution-inductively coupled plasma mass spectrometry (ID-ICP-MS) was applied to the determination of lead in natural water, and its precision was compared to that of the conventional calibration method. Trace amount of lead in natural water was preconcentrated using a chelating resin of Chelex 100 before ID-ICP-MS, where the preconcentration factor of 50 was obtained. The concentrations of lead were 26.1 pg/g and 266 pg/g for the certified river reference materials (JAC 0031) and the pond water sample, respectively, with the RSD less than 1%.
引用
收藏
页码:7 / 10
页数:4
相关论文
共 10 条
[1]   DETERMINATION OF TRACE-METALS IN A RIVER WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1987, 59 (05) :778-783
[2]   APPLICATION OF ISOTOPE-DILUTION ANALYSIS INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY TO THE PRECISE DETERMINATION OF SILVER AND ANTIMONY IN PURE COPPER [J].
CHIBA, K ;
INAMOTO, I ;
SAEKI, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (02) :115-119
[3]  
Date A. R., 1988, APPL INDUCTIVELY COU
[4]  
HIRAO Y, 1974, BUNSEKI KAGAKU, V23, P158
[5]  
Jarvis K. E., 1992, HDB INDUCTIVELY COUP, DOI [DOI 10.1007/BF02290251, 10.1007/BF02290251]
[6]   Determination of trace impurities on silicon-wafer surface by isotope dilution analysis using electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Komoda, M ;
Chiba, K ;
Uchida, H .
ANALYTICAL SCIENCES, 1996, 12 (01) :21-25
[7]   DETERMINATION OF GALLIUM IN PURE ALUMINUM BY ISOTOPE-DILUTION ANALYSIS WITH SOLVENT-EXTRACTION USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
MAKISHIMA, A ;
INAMOTO, I ;
CHIBA, K .
APPLIED SPECTROSCOPY, 1990, 44 (01) :91-95
[8]   DETERMINATION OF LEAD ISOTOPE RATIOS AND CONCENTRATIONS IN SEA-WATER BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY AFTER PRECONCENTRATION USING CHELEX-100 [J].
MIYAZAKI, A ;
REIMER, RA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (03) :449-452
[9]   HIGH ACCURACY DETERMINATION OF CALCIUM IN BLOOD SERUM BY ISOTOPE DILUTION MASS-SPECTROMETRY [J].
MOORE, LJ ;
MACHLAN, LA .
ANALYTICAL CHEMISTRY, 1972, 44 (14) :2291-&
[10]  
STURGEON RE, 1986, CAN J CHEM, V65, P961