Monte Carlo and experimental derivation of TG43 dosimetric parameters for CSM-type Cs-137 sources

被引:33
作者
Pérez-Calatayud, J
Granero, D
Casal, E
Ballester, F
Puchades, V
机构
[1] Univ Valencia, CSIC, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[2] Univ Valencia, CSIC, IFIC, E-46100 Burjassot, Spain
[3] Grp IMO SFA, Madrid, Spain
关键词
Cs-137; dosimetry; LDR brachytherapy; Monte Carlo; intracavitary sources;
D O I
10.1118/1.1834835
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study, complete dosimetric datasets for the CSM2 and CSM3 Cs-137 sources were obtained using the Monte Carlo GEANT4 code. The application of this calculation method was experimentally validated with thermoluminescent dosimetry (TLD). Functions and parameters following the TG43 formalism are presented: the dose rate constant, the radial dose functional, and the anisotropy function. In addition, to aid the quality control process on treatment planning systems, a two-dimensional (213) rectangular dose rate table (the traditional along-away table), coherent with the TG43 dose calculation formalism, is given. The data given in this study complement existing information for both sources on the following aspects: (i) the source asymmetries were considered explicitly in the Monte Carlo calculations, (ii) TG43 data were derived directly from Monte Carlo calculations, (iii) the radial range of the different tables was increased as well as the angular resolution in the anisotropy function, including angles close to the longitudinal source axis. The CSM2 source TG-43 data of Liu et al. [Med. Phys. 31, 477-483 (2004)] are not consistent with the Williamson 2D along-away data [Int. J. Radiat. Oncol., Biol., Phys. 15, 227-237 (1988)] at distances closer than approximately 2 cm from the source. The data presented here for this source are consistent with this 2D along-away table, and are suitable for use in clinical practice. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:28 / 36
页数:9
相关论文
共 20 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   A Monte Carlo investigation of the dosimetric characteristics of the CSM11 137Cs source from CIS [J].
Ballester, F ;
Lluch, JL ;
Limami, Y ;
Serrano, MA ;
Casal, E ;
Pérez-Calatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 2000, 27 (09) :2182-2189
[3]   Monte Carlo calculation of dose rate distributions around Ir-192 wires [J].
Ballester, F ;
Hernandez, C ;
PerezCalatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 1997, 24 (08) :1221-1228
[4]   Monte Carlo calculations of dose rate distributions around the Amersham CDCS-M-type 137Cs source [J].
Casal, E ;
Ballester, F ;
Lluch, JL ;
Pérez-Calatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 2000, 27 (01) :132-140
[5]   Determination of 137Cs dosimetry parameters according to the AAPM TG-43 formalism [J].
Liu, LZ ;
Prasad, SC ;
Bassano, DA .
MEDICAL PHYSICS, 2004, 31 (03) :477-483
[6]   EFFECTIVE ATTENUATION IN WATER OF GAMMA RAYS OF GOLD 198 IRIDIUM 192 CESIUM 137 RADIUM 226 AND COBALT 60 [J].
MEISBERGER, LL ;
KELLER, RJ ;
SHALEK, RJ .
RADIOLOGY, 1968, 90 (05) :953-+
[7]  
Nath R, 1999, Cardiovasc Radiat Med, V1, P72, DOI 10.1016/S1522-1865(98)00007-9
[8]   Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56 [J].
Nath, R ;
Anderson, LL ;
Meli, JA ;
Olch, AJ ;
Stitt, JA ;
Williamson, JF .
MEDICAL PHYSICS, 1997, 24 (10) :1557-1598
[9]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[10]   Phantom size in brachytherapy source dosimetric studies [J].
Pérez-Calatayud, J ;
Granero, D ;
Ballester, F .
MEDICAL PHYSICS, 2004, 31 (07) :2075-2081