The calibration of CT Hounsfield units for radiotherapy treatment planning

被引:830
作者
Schneider, U
Pedroni, E
Lomax, A
机构
[1] UNIV MUNICH,SECT PHYS,MED PHYS GRP,GARCHING,GERMANY
[2] PAUL SCHERRER INST,DEPT RADIAT MED,VILLIGEN,SWITZERLAND
关键词
D O I
10.1088/0031-9155/41/1/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Computer tomographic (CT) scans are used to correct for tissue inhomogeneities in radiotherapy treatment planning. in order to guarantee a precise treatment, it is important to obtain the relationship between CT Hounsfield units and electron densities (or proton stopping powers for proton radiotherapy), which is the basic input for radiotherapy planning systems which consider tissue heterogeneities. A method is described to determine improved CT calibrations for biological tissue (a stoichiometric calibration) based on measurements using tissue equivalent materials. The precision of this stoichiometric calibration and the more usual tissue substitute calibration is determined by a comparison of calculated proton radiographic images based on these calibrations and measured radiographs of a biological sample. It has been found that the stoichiometric calibration is more precise than the tissue substitute calibration.
引用
收藏
页码:111 / 124
页数:14
相关论文
共 15 条
[1]  
[Anonymous], ICRP PUBLICATION
[2]   COMPUTED-TOMOGRAPHY FOR RADIOTHERAPY PLANNING [J].
BATTISTA, JJ ;
RIDER, WD ;
VANDYK, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1980, 6 (01) :99-107
[3]  
BICHSEL H, 1972, AM I PHYSICS HDB
[4]   TREATMENT PLANNING FOR HEAVY-ION RADIOTHERAPY [J].
CHEN, GTY ;
SINGH, RP ;
CASTRO, JR ;
LYMAN, JT ;
QUIVEY, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (10) :1809-1819
[5]   AN ELECTRON-DENSITY CALIBRATION PHANTOM FOR CT-BASED TREATMENT PLANNING COMPUTERS [J].
CONSTANTINOU, C ;
HARRINGTON, JC ;
DEWERD, LA .
MEDICAL PHYSICS, 1992, 19 (02) :325-327
[6]  
CONSTANTINOU C, 1974, THESIS LONDON U
[7]   X-RAY ATTENUATION COEFFICIENTS OF ELEMENTS AND MIXTURES [J].
JACKSON, DF ;
HAWKES, DJ .
PHYSICS REPORTS-REVIEW SECTION OF PHYSICS LETTERS, 1981, 70 (03) :169-233
[8]  
Janni J F, 1982, ATOM DATA NUCL DATA, V27, P212
[9]   ACCEPTANCE TESTING COMPUTERIZED RADIATION-THERAPY TREATMENT PLANNING SYSTEMS - DIRECT UTILIZATION OF CT SCAN DATA [J].
MCCULLOUGH, EC ;
HOLMES, TW .
MEDICAL PHYSICS, 1985, 12 (02) :237-242
[10]  
MOYERS M, 1993, COMMUNICATION