Suture anchor strength revisited

被引:71
作者
Barber, FA
Herbert, MA
Click, JN
机构
[1] PLANO ORTHOPED & SPORTS MED CTR,PLANO,TX
[2] MED CITY DALLAS HOSP,ADV SURG INST,DALLAS,TX
来源
ARTHROSCOPY | 1996年 / 12卷 / 01期
关键词
suture anchor; Mitek; miniHarpoon; PeBA; ROC; miniMitek; Fastak; ultimate load;
D O I
10.1016/S0749-8063(96)90216-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The rapid proliferation of suture anchors continues. Our prior report on the pullout strength of 14 different anchors is supplemented by a similar test conducted on 8 additional anchors. Comparative data on modes of failure and failure strengths (ultimate loads to failure) for these new devices are compared statistically with the previously tested anchors. In a fresh never-frozen porcine femur model, 10 samples of each of the additional anchors tested were threaded with stainless steel sutures and inserted into three different test areas (diaphyseal cortex, metaphyseal cortex, and a cancellous trough). Tensile stress parallel to the axis of insertion was applied at a rate of 12.5 mm/s by an Instron 1321 testing machine (Instron Corp, Canton, MA) until failure and mean anchor failure strengths calculated. The anchors tested were the Mitek G2 as a control, miniMitek, Mitek Superanchor, Mitek Rotator Cuff anchor (Mitek Products, Westwood, MA), Innovasive Devices Radial Osteal Compression device (Innovasive Devices, Hopkinton, MA), Arthrex Fastak (Arthrex Inc, Naples, FL), Arthrotek miniHarpoon (Arthrotek, Warsaw, IN), Orthopedic Biosystems PeBA 3 and PeBA 5 (Orthopedic Biosystems, Scottsdale, AZ), and AME 5.5 screw (American Medical Electronics, Richardson, TX). Failure mode (anchor pullout, suture eyelet cut out, or wire breakage) was generally consistent for each anchor type. The size of insertion hole is clinically important and each anchor's performance was evaluated as a function of its minor diameter or drill hole. For screw anchors, the larger the minor diameter of the screw, the higher the mean failure strengths in all three test areas (P = .001). However, larger drill holes for non-screw anchors resulted in lower mean failure strengths in cancellous bone (P = .03) and diaphyseal cortex (P < .005).
引用
收藏
页码:32 / 38
页数:7
相关论文
共 23 条
[11]   USE OF THE SUTURE ANCHOR IN OPEN BANKART RECONSTRUCTION - A FOLLOW-UP REPORT [J].
LEVINE, WN ;
RICHMOND, JC ;
DONALDSON, WR .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1994, 22 (05) :723-726
[12]  
Norlin R, 1994, J Shoulder Elbow Surg, V3, P381, DOI 10.1016/S1058-2746(09)80023-2
[13]  
Paulos L E, 1993, J Shoulder Elbow Surg, V2, P305, DOI 10.1016/1058-2746(93)90076-S
[14]   ARTHROSCOPICALLY ENHANCED MINIAPPROACH TO ROTATOR CUFF REPAIR [J].
PAULOS, LE ;
KODY, MH .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1994, 22 (01) :19-25
[15]  
PEDERSON B, 1991, Journal of Foot Surgery, V30, P48
[16]   THE MITEK BONE ANCHOR - APPLICATION TO THE HAND, WRIST AND ELBOW [J].
REHAK, DC ;
SOTEREANOS, DG ;
BOWMAN, MW ;
HERNDON, JH .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1994, 19A (05) :853-860
[17]   MODIFICATION OF THE BANKART RECONSTRUCTION WITH A SUTURE ANCHOR - REPORT OF A NEW TECHNIQUE [J].
RICHMOND, JC ;
DONALDSON, WR ;
FU, F ;
HARNER, CD .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1991, 19 (04) :343-346
[18]  
Saitoh S, 1994, J Shoulder Elbow Surg, V3, P234, DOI 10.1016/S1058-2746(09)80041-4
[19]   RECONSTRUCTION FOR SUBTALAR INSTABILITY - A REVIEW [J].
SCHON, LC ;
CLANTON, TO ;
BAXTER, DE .
FOOT & ANKLE, 1991, 11 (05) :319-325
[20]   SOFT-TISSUE RECONSTRUCTION IN THE SHOULDER - COMPARISON OF SUTURE ANCHORS, ABSORBABLE STAPLES, AND ABSORBABLE TACKS [J].
SHALL, LM ;
CAWLEY, PW .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1994, 22 (05) :715-718