Rotating magnetic macrospheres as heating mechanism for remote controlled drug release

被引:13
作者
Steinke, Franziska
Andrae, Wilfried
Heide, Rainer
Werner, Christoph
Bellemann, Matthias Erich
机构
[1] Univ Appl Sci, Dept Biomed Engn, Jena, Germany
[2] Inst Phys HochTechnol EV, Jena, Germany
关键词
friction losses; rotating magnetic field; special torque-meter; thermal resistance; viscosity;
D O I
10.1016/j.jmmm.2006.10.1193
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A permanent magnetic macrosphere (diameter: 5 mm) spherically seated in an oil bearing inside an experimental capsule (comparable to a hard gelatine capsule size 2) is turned by a rotating magnetic field (H <= 5 kA/m; frequency v <= 500 Hz) and causes a temperature rise up to about 60 degrees C. In order to find further possible improvements, the experimental results were compared to theoretical expectations. First experiments using improved thermal isolation yielded temperatures of about 100 degrees C. The heating can be used as a mechanism to remotely release drugs in the gastrointestinal tract. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 218
页数:3
相关论文
共 7 条
[1]   A novel magnetic method for examination of bowel motility [J].
Andrä, W ;
Danan, H ;
Eitner, K ;
Hocke, M ;
Kramer, HH ;
Parusel, H ;
Saupe, P ;
Werner, C ;
Bellemann, ME .
MEDICAL PHYSICS, 2005, 32 (09) :2942-2944
[2]  
ANDRA W, 2004, Patent No. 10310825
[3]  
DUTZ S, 2005, BIOMED TECH S, V50, P601
[4]  
KNAUFT S, 2004, BIOMED TECHN S2, V49, P724
[5]  
LANDAU LD, 1978, LEHRBUCH THEORETISCH, V6
[6]  
LEOPOLD CS, 2001, DRUG TARGETING ORGAN
[7]   Remote controlled capsules in human drug absorption (HDA) studies [J].
Wilding, IR ;
Prior, DV .
CRITICAL REVIEWS IN THERAPEUTIC DRUG CARRIER SYSTEMS, 2003, 20 (06) :405-431