Transit time flow measurement: Experimental validation and comparison of three different systems

被引:56
作者
Beldi, G
Bosshard, A
Hess, OM
Althaus, U
Walpoth, BH [1 ]
机构
[1] Univ Hosp Insel, Dept Cardiovasc Surg, CH-3010 Bern, Switzerland
[2] Univ Hosp Insel, Div Cardiol, CH-3010 Bern, Switzerland
关键词
D O I
10.1016/S0003-4975(00)01246-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Bloodflow measurements are of major clinical importance for quality control in vascular surgery. They allow detection of low-flow situations which may influence outcome adversely. The purpose of the present study was to validate three different flow systems for measuring absolute blood flow. Methods. Measurements were performed in an experimental flow model using arteries and veins and blood or saline at two different temperatures. As a reference method true flow was measured by volume sampling. Results. Correlation coefficients between transit time flow and true flow measurements ranged between 0.71 and 0.92. Systematic overestimation and underestimation of transit time flow were observed, but after second-order correction all correlations were excellent, ranging from 0.93 to 0.95 irrespective of flow medium and temperature. Conclusions. Transit time flow measurements are exact and reproducible. Second-order correction yields good accuracy and high precision, with minimal differences among the three systems evaluated. (Ann Thorac Surg 2000;70:212-7) (C) 2000 by The Society of Thoracic Surgeons.
引用
收藏
页码:212 / 217
页数:6
相关论文
共 20 条
[1]  
[Anonymous], 1978, P SAN DIEG BIOM S
[2]  
ASCER E, 1984, J VASC SURG, V1, P817
[3]   OUTFLOW RESISTANCE MEASUREMENT DURING INFRAINGUINAL ARTERIAL RECONSTRUCTIONS - A RELIABLE PREDICTOR OF LIMB SALVAGE [J].
ASCER, E ;
WHITE, SA ;
VEITH, FJ ;
MORIN, L ;
FREEMAN, K ;
GUPTA, SK .
AMERICAN JOURNAL OF SURGERY, 1987, 154 (02) :185-188
[4]   INTRAOPERATIVE OUTFLOW RESISTANCE AS A PREDICTOR OF LATE PATENCY OF FEMOROPOPLITEAL AND INFRAPOPLITEAL ARTERIAL BYPASSES [J].
ASCER, E ;
VEITH, FJ ;
WHITEFLORES, SA ;
MORIN, L ;
GUPTA, SK ;
LESSER, ML .
JOURNAL OF VASCULAR SURGERY, 1987, 5 (06) :820-827
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Burton RG, 1984, MED ELECTRONICS, V15, P68
[7]  
HARTMAN JC, 1994, J PHARMACOL TOXICOL, V31, P153
[8]   Role of graft flow measurement technique in anastomotic quality assessment in minimally invasive CABG [J].
Jaber, SF ;
Koenig, SC ;
BhaskerRao, B ;
VanHimbergen, DJ ;
Cerrito, PB ;
Ewert, DJ ;
Gray, LA ;
Spence, PA .
ANNALS OF THORACIC SURGERY, 1998, 66 (03) :1087-1092
[9]  
KOLIN A, 1975, LIFE SCI, V16, P501
[10]   Validation of a new transit time ultrasound flowmeter in man [J].
Laustsen, J ;
Pedersen, EM ;
Terp, K ;
Steinbruchel, D ;
Kure, HH ;
Paulsen, PK ;
Jorgensen, H ;
Paaske, WP .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1996, 12 (01) :91-96