Limitations of current validation protocols for home blood pressure monitors for individual patients

被引:56
作者
Gerin, W
Schwartz, AR
Schwartz, JE
Pickering, TG
Davidson, KW
Bress, J
O'Brien, E
Atkins, N
机构
[1] Mt Sinai Sch Med, Zena & Michael A Wiener Cardiovasc Inst, New York, NY 10029 USA
[2] Yale Univ, Sch Med, New Haven, CT 06520 USA
[3] SUNY Stony Brook, Stony Brook, NY 11794 USA
[4] Cornell Univ, Joan & Sanford I Weill Med Coll, New York, NY USA
[5] Beaumont Hosp, Blood Pressure Unit, Dublin 9, Ireland
关键词
hypertension; validation; measurement; blood pressure; blood pressure monitoring;
D O I
10.1097/00126097-200212000-00004
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background Automatic blood pressure monitoring conducted at home is increasingly used in the diagnosis and management of hypertension. We assessed the adequacy of existing British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) validation standards for automatic blood pressure monitoring devices. Subject and methods A theoretical study and an empirical test are presented to estimate the proportion of persons for whom a blood pressure monitor validated according to existing BHS and AAMI standards would inaccurate. Results The results suggest that a major limitation of both protocols is the lack of attention given to the number of individual patients for whom a monitor may e inaccurate. A blood pressure monitor that meets the AAMI and BHS validation criteria may report blood pressures in error by more than 5 mmHg for more than half of the people. Conclusions A validation standard that does not take account of the person-effects on error will lead to a substantial proportion of persons using self-monitors that are systematically inaccurate for that person. (C) 2002 Lippincott Williams Wilkins.
引用
收藏
页码:313 / 318
页数:6
相关论文
共 26 条
  • [1] [Anonymous], 1997, ARCH INTERN MED, V157, P2413, DOI [10.1001/archinte.1997.00440420033005, DOI 10.1001/ARCHINTE.1997.00440420033005]
  • [2] *ASS ADV MED INSTR, 1987, AM STAND EL AUT SPHY
  • [3] Association for the Advancement of Medical Instrumentation, 1993, AM NAT STAND EL AUT
  • [4] Chalmers J, 1999, J HYPERTENS, V17, P151
  • [5] THE INFLUENCE OF AMBULATORY BLOOD-PRESSURE MONITORING ON THE DESIGN AND INTERPRETATION OF TRIALS IN HYPERTENSION
    COATS, AJS
    RADAELLI, A
    CLARK, SJ
    CONWAY, J
    SLEIGHT, P
    [J]. JOURNAL OF HYPERTENSION, 1992, 10 (04) : 385 - 391
  • [6] GERIN W, 1999, CURR HYPERTENS REP, V1, P1
  • [7] Mercury sphygmomanometers should not be abandoned: An advisory statement from the council for high blood pressure research, American Heart Association
    Jones, DW
    Frohlich, ED
    Grim, CM
    Grim, CE
    Taubert, KA
    [J]. HYPERTENSION, 2001, 37 (02) : 185 - 186
  • [8] Multiple clinic and home blood pressure measurements versus ambulatory blood pressure monitoring
    Jula, A
    Puukka, P
    Karanko, H
    [J]. HYPERTENSION, 1999, 34 (02) : 261 - 266
  • [9] Mancia G, 1997, CIRCULATION, V95, P1464
  • [10] Mee F, 1998, Blood Press Monit, V3, P353