Lithium dilution cardiac output measurement: A comparison with thermodilution

被引:102
作者
Linton, R
Band, D
OBrien, T
Jonas, M
Leach, R
机构
[1] JOHN RADCLIFFE HOSP,NUFFIELD DEPT ANAESTHET,OXFORD OX3 9DU,ENGLAND
[2] ST THOMAS HOSP,INTENS CARE UNIT,LONDON,ENGLAND
关键词
cardiac output; indicator dilution; thermodilution; critical care; hemodynamics; heart; pulmonary artery catheterization; monitoring; physiologic; lithium;
D O I
10.1097/00003246-199711000-00015
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare the results of cardiac output measurements obtained by lithium dilution and thermodilution. Design: Case series, observational study. Setting: High dependency postoperative unit and intensive care unit of a teaching hospital. Patients: Forty patients were studied, Thirty-four patients had undergone heart surgery requiring cardiopulmonary bypass within the previous 2 days; the diagnoses in the other patients were myocardial infarct (n = 2), septicemia (n = 2), adult respiratory distress syndrome, and pericardectomy. Interventions: Cardiac output was measured five times in each patient, using lithium dilution (single measurement) and bolus thermodilution (series of three to six measurements according to standard clinical practice, taking the average of the closest three), In a subgroup of 14 patients, cardiac output was also measured using ''continuous thermodilution.'' Measurements and Main Results: Comparing lithium dilution with bolus thermodilution, the mean of the differences (lithium dilution - thermodilution) was -0.25 +/- 0.46 [so] L/min, Linear regression analysis gave y = 0.31 + 0.89x (r(2) = .94) for lithium dilution vs, thermodilution. Conclusions: The overall agreement between the two methods was good, The variability of the thermodilution measurements was greater than that of the lithium dilution measurements, The lithium dilution method is at least as accurate as bolus thermodilution and, since pulmonary artery catheterization is not needed, it has the advantages of being safe and quick to perform.
引用
收藏
页码:1796 / 1800
页数:5
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   A PROSPECTIVE-STUDY OF COMPLICATIONS OF PULMONARY-ARTERY CATHETERIZATIONS IN 500 CONSECUTIVE PATIENTS [J].
BOYD, KD ;
THOMAS, SJ ;
GOLD, J ;
BOYD, AD .
CHEST, 1983, 84 (03) :245-249
[3]   HEMODYNAMIC MONITORING - INVASIVE TECHNIQUES [J].
BUCHBINDER, N ;
GANZ, W .
ANESTHESIOLOGY, 1976, 45 (02) :146-155
[4]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[5]   Studies on the circulation IV. Further analysis of the injection method, and of changes in hemodynamtcs under physiological and pathological conditions [J].
Hamilton, WF ;
Moore, JW ;
Kinsman, JM ;
Spurling, RG .
AMERICAN JOURNAL OF PHYSIOLOGY, 1932, 99 (03) :534-551
[6]  
HORST HM, 1984, SURG GYNECOL OBSTET, V159, P229
[7]   CONTINUOUS CARDIAC-OUTPUT MEASUREMENTS IN THE PERIOPERATIVE PERIOD [J].
JAKOBSEN, CJ ;
MELSEN, NC ;
ANDRESEN, EB .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 (04) :485-488
[8]  
KHALIL HH, 1963, LANCET, V1, P1352
[9]   CARDIAC-OUTPUT MEASUREMENT IN CRITICALLY ILL PATIENTS - COMPARISON OF CONTINUOUS AND CONVENTIONAL THERMODILUTION TECHNIQUES [J].
LEFRANT, JY ;
BRUELLE, P ;
RIPART, J ;
IBANEZ, F ;
AYA, G ;
PERAY, P ;
SAISSI, G ;
DELACOUSSAYE, JE ;
ELEDJAM, JJ .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1995, 42 (11) :972-976
[10]   A NEW METHOD OF MEASURING CARDIAC-OUTPUT IN MAN USING LITHIUM DILUTION [J].
LINTON, RAF ;
BAND, DM ;
HAIRE, KM .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (02) :262-266