A novel type of a miniature transducer for potentiometric sensors was designed and tested. The transducer is made from an ordinary printed circuit board using a typical technology. As a result, a back-side contact structure was obtained, which can be mounted in flow systems. The advantage of the fabrication process is its cheapness and flexibility in designing modified structures. The measurement properties of the transducer were tested as a chloride sensor. Then the structure was covered with polyHEMA and a potassium-selective membrane and tested as a potassium sensor. The performance and the long-term stability of the sensors with plasticized PVC and Siloprene based potassium-selective membranes were compared. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:103 / 109
页数:7
相关论文
共 13 条
[1]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[2]   Thirty years of ISFETOLOGY - What happened in the past 30 years and what may happen in the next 30 years [J].
Bergveld, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2003, 88 (01) :1-20
[3]   MICROFABRICATION TECHNOLOGY OF FLEXIBLE MEMBRANE-BASED SENSORS FOR IN-VIVO APPLICATIONS [J].
BUCK, RP ;
COSOFRET, VV ;
LINDNER, E ;
UFER, S ;
MADARAS, MB ;
JOHNSON, TA ;
ASH, RB ;
NEUMAN, MR .
ELECTROANALYSIS, 1995, 7 (09) :846-851
[4]   Novel head for testing and measurement of chemical microsensors [J].
Chudy, M ;
Dybko, A ;
Wróblewski, W ;
Brzózka, Z .
ANALYTICA CHIMICA ACTA, 2001, 429 (02) :347-355
[5]  
DYBKO A, 2001, METROLOGY MEASURING, V8, P263
[6]   Solvent-resistant carbon electrodes screen printed onto plastic for use in biosensors [J].
Kroger, S ;
Turner, APF .
ANALYTICA CHIMICA ACTA, 1997, 347 (1-2) :9-18
[7]  
LYSKO JM, 2000, Patent No. 339473
[8]   Technological aspects of potentiometric BSC-type microsensor fabrication [J].
Pijanowska, DG ;
Wygldacz, K ;
Jazwinski, J ;
Lysko, JM ;
Koszur, J ;
Brzozka, Z ;
Malinowska, E .
OPTOELECTRONIC AND ELCTRONIC SENSORS IV, 2001, 4516 :32-35
[9]   QUALITY CLASSIFICATION OF GRAIN USING A SENSOR ARRAY AND PATTERN-RECOGNITION [J].
STETTER, JR ;
FINDLAY, MW ;
SCHROEDER, KM ;
YUE, C ;
PENROSE, WR .
ANALYTICA CHIMICA ACTA, 1993, 284 (01) :1-11
[10]   MODIFICATION OF ISFETS BY COVALENT ANCHORING OF POLY(HYDROXYETHYL METHACRYLATE) HYDROGEL - INTRODUCTION OF A THERMODYNAMICALLY DEFINED SEMICONDUCTOR-SENSING MEMBRANE INTERFACE [J].
SUDHOLTER, EJR ;
VANDERWAL, PD ;
SKOWRONSKAPTASINSKA, M ;
VANDENBERG, A ;
BERGVELD, P ;
REINHOUDT, DN .
ANALYTICA CHIMICA ACTA, 1990, 230 (01) :59-65