Postmortem analysis of consecutively retrieved asymmetric porous-coated tibial components

被引:27
作者
Bloebaum, RD [1 ]
Bachus, KN [1 ]
Jensen, JW [1 ]
Hofmann, AA [1 ]
机构
[1] VA Med Ctr, Bone & Joint Res Lab 151F, Salt Lake City, UT 84148 USA
关键词
backscattered electron; bone ingrowth; porous coating; tibia; total knee;
D O I
10.1016/S0883-5403(97)90162-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The objective of this investigation was to conduct a postmortem analysis of 8 porous-coated asymmetric tibial components to measure the extent of radiolucencies and bone ingrowth. With the use of radiographic, electron microscope, and histologic analysis techniques, a quantitative postmortem study of 8 consecutively retrieved porous-coated tibial components was conducted. Time in situ averaged 47 +/- 36 months. The components were secured with 4 pegs and 2 screws. Autograft bone chips were applied to the resected tibia during implantation. Contact radiographs of an average of 8 3-mm sections from each implant revealed that 73% +/- 17% of the porous coating had no apparent radiolucencies present between the host bone and porous coating for the series. Backscattered electron imaging showed that the bone ingrowth averaged 6% +/- 2%. Histologic analysis was unable to demonstrate any adverse cellular response. The analysis suggested that this asymmetric implant design is stable and biocompatible and has potential for long-term clinical durability.
引用
收藏
页码:920 / 929
页数:10
相关论文
共 34 条
[1]  
BACHUS KN, 1992, CELL MATER, V2, P13
[2]   TOTAL KNEE ARTHROPLASTY IN THE MANAGEMENT OF PROXIMAL TIBIAL STRESS-FRACTURES [J].
TOMLINSON, MPW ;
DINGWALL, IM ;
PHILLIPS, H .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :707-713
[3]  
BLOEBAUM RD, 1994, CLIN ORTHOP RELAT R, P2
[4]   Postmortem analysis of bone growth into porous-coated acetabular components [J].
Bloebaum, RD ;
Mihalopoulus, NL ;
Jensen, JW ;
Dorr, LD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (07) :1013-1022
[5]  
BLOEBAUM RD, 1989, J HISTOTECHNOL, V12, P307
[6]   MINERAL APPOSITION RATES OF HUMAN CANCELLOUS BONE AT THE INTERFACE OF POROUS-COATED IMPLANTS [J].
BLOEBAUM, RD ;
BACHUS, KN ;
MOMBERGER, NG ;
HOFMANN, AA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (05) :537-544
[7]  
BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, P179
[8]  
BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, P97
[9]  
COOK SD, 1992, CLIN ORTHOP RELAT R, P163
[10]  
COOK SD, 1988, CLIN ORTHOP RELAT R, P90