Poly(tetrafluoroethylene) film housing of carbon fibers using capillary-pull technology for one-stage fabrication of carbon disk ultramicroelectrodes and their characterization

被引:22
作者
Zhang, XJ [1 ]
Ogorevc, B [1 ]
机构
[1] Natl Inst Chem, Analyt Chem Lab, Ljubljana 1001, Slovenia
关键词
D O I
10.1021/ac970833g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fabrication and characterization of carbon disk ultramicroelectrodes (CDUMEs), embodied in a pulled Teflon capillary, with overall tip dimension of similar to 10 mu m in diameter, are described. A CDUME was constructed by inserting a carbon fiber 7 mu m in diameter into a commercial Teflon capillary, which was followed by pulling the capillary by means of a microelectrode puller employing appropriate heating and timing, to produce a self-sealing thin Tenon film insulation coating. Then, the so-coated carbon fiber was cut to expose a fresh carbon fiber disk, The proposed one-stage preparation method is fast (similar to 5 min), very simple, and inexpensive and eliminates the need for separate embodying, insulation, and sealing steps. It results in CDUMEs exhibiting excellent electrochemical behavior. Scanning electron and optical microscopy, voltammetry, and amperometry were employed to characterize these electrodes. Cyclic voltammograms of ferricyanide in aqueous media and of ferrocene in acetonitrile media displayed low-noise, low-background, sigmoidal responses with virtually no current hysteresis. To check the analytical applicability of these electrodes, a testing with adrenaline was performed by applying the differential pulse voltammetry mode. A linear calibration over the concentration range from 2.5 x 10(-6) to 5.0 x 10(-4) mol/L in pH 7.2 phosphate buffer solution was obtained with a detection limit of 7.0 x 10(-7) mol/L. The proposed CDUMEs with flexible and nonfragile Tenon housing exhibit very low electrical noise and can be reproduced multiple times by simply recutting the tip.
引用
收藏
页码:1646 / 1651
页数:6
相关论文
共 29 条
  • [11] ELECTROCHEMICAL DETECTION IN MICROCOLUMN SEPARATIONS
    EWING, AG
    MESAROS, JM
    GAVIN, PF
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (09) : A527 - A537
  • [12] ANALYTICAL-CHEMISTRY IN MICROENVIRONMENTS - SINGLE NERVE-CELLS
    EWING, AG
    STREIN, TG
    LAU, YY
    [J]. ACCOUNTS OF CHEMICAL RESEARCH, 1992, 25 (10) : 440 - 447
  • [13] ETCHED CARBON-FIBER ELECTRODES AS AMPEROMETRIC DETECTORS OF CATECHOLAMINE SECRETION FROM ISOLATED BIOLOGICAL CELLS
    KAWAGOE, KT
    JANKOWSKI, JA
    WIGHTMAN, RM
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (15) : 1589 - 1594
  • [14] KELLY BS, 1986, ANAL CHIM ACTA, V186, P79
  • [15] MICROELECTRODES FOR ANODIC-STRIPPING VOLTAMMETRY PREPARED BY HEAT SEALING THIN FIBERS OR WIRES IN A POLYPROPYLENE MATRIX
    NYHOLM, L
    WIKMARK, G
    [J]. ANALYTICA CHIMICA ACTA, 1992, 257 (01) : 7 - 13
  • [16] CONSTRUCTION OF SUBMICROMETER VOLTAMMETRIC ELECTRODES
    PENDLEY, BD
    ABRUNA, HD
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (07) : 782 - 784
  • [17] FABRICATION AND USE OF NANOMETER-SIZED ELECTRODES IN ELECTROCHEMISTRY
    PENNER, RM
    HEBEN, MJ
    LONGIN, TL
    LEWIS, NS
    [J]. SCIENCE, 1990, 250 (4984) : 1118 - 1121
  • [18] ELECTROCHEMICALLY PREPARED INSULATION FOR CARBON-FIBER MICROELECTRODES
    POTJEKAMLOTH, K
    JANATA, J
    JOSOWICZ, M
    [J]. BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1989, 93 (12): : 1480 - 1485
  • [19] CHARACTERIZATION OF THE DIFFUSION LIMITED CURRENT AT ULTRA-SMALL CARBON RING ELECTRODES
    SARACENO, RA
    EWING, AG
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 257 (1-2): : 83 - 93
  • [20] A simple method for insulating carbon fiber microelectrodes using anodic electrophoretic deposition of paint
    Schulte, A
    Chow, RH
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (17) : 3054 - 3058