Ultrasound evaluation of endovascular repair of abdominal aortic aneurysms

被引:11
作者
Kronzon, I
Tunick, PA
Rosen, R
Riles, T
机构
[1] NYU, Med Ctr, Dept Med, New York, NY 10016 USA
[2] NYU, Med Ctr, Dept Radiol, New York, NY 10016 USA
[3] NYU, Med Ctr, Dept Surg, New York, NY 10016 USA
关键词
D O I
10.1016/S0894-7317(98)70106-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Endovascular repair of an abdominal aortic aneurysm (AAA) offers a minimally invasive alternative to an open surgical procedure in selected patients. The purpose of this study was to examine the usefulness of ultrasonography for evaluating the results of endovascular repair. Methods. We studied 17 patients who underwent endovascular repair. In 10 patients a bifurcated prosthesis was positioned below the renal arteries with the bifurcated branches in the iliac arteries. The other 7 patients had a nonbifurcated infrarenal prosthesis. Results: In each patient the AAA and the entire prosthesis, including its bifurcated branches, were visualized. The mean AAA diameter was 5.0 +/- 0.6 cm. The mean prosthesis body diameter was 2.2 +/- 0.3 cm, and the diameters of the bifurcated limbs were 1.0 to 1.2 cm. Color Doppler studies revealed blood flow limited to the prosthetic lumen and its bifurcation in 16 patients; the space between the prosthesis and the AAA wall was clotted in these patients. In 1 patient a communication was seen between the prosthesis and the AAA lumen through a dehiscence in the distal attachment. Conclusion: Ultrasonography is a simple, noninvasive tool for the evaluation of the results of endovascular repair of AAA and can detect complications of this procedure.
引用
收藏
页码:377 / 380
页数:4
相关论文
共 9 条
[1]  
HERTZER NR, 1984, J VASC SURG, V1, P13
[2]   NONRUPTURED ABDOMINAL AORTIC-ANEURYSM - 6-YEAR FOLLOW-UP RESULTS FROM THE MULTICENTER PROSPECTIVE CANADIAN ANEURYSM STUDY [J].
JOHNSTON, KW ;
KALMAN, PG ;
AMELI, FM ;
EVANS, G ;
MCPHAIL, NV ;
SLADEN, JG ;
AUSTERWEIL, A ;
AU, HH ;
BAIRD, RJ ;
BALACHANDRA, VK ;
BARBER, GG ;
BASIAN, H ;
BLUNDELL, PE ;
BROWN, PM ;
BRUNEAU, L ;
CALLAGHAN, JC ;
CARROLL, SE ;
CHIU, CJ ;
DEROSE, G ;
DOOBAY, BS ;
DOUVILLE, Y ;
DOWNS, AR ;
DUNN, RS ;
DUTTON, JW ;
ERRET, LE ;
FONG, HMT ;
FRY, PD ;
GELFAND, ET ;
GOLDBACH, MM ;
GOLDBERG, MR ;
GOLDSTEIN, AS ;
GOODING, JT ;
GRANT, KC ;
GUNSTENSEN, J ;
HARRIS, KA ;
HEIDE, AR ;
HILDEBRAND, H ;
HUNTING, MC ;
HYATT, HJ ;
JAMIESON, WG ;
KING, WL ;
LAURENDEAU, F ;
LITHERLAND, HK ;
MACKEAN, GL ;
MADDISON, GD ;
MAGGISANO, R ;
MEADS, GE ;
MICHALSKI, AH ;
MODREY, DL ;
MOFFAT, RC .
JOURNAL OF VASCULAR SURGERY, 1994, 20 (02) :163-170
[3]   OPERATIVE MORTALITY-RATES FOR INTACT AND RUPTURED ABDOMINAL AORTIC-ANEURYSMS IN MICHIGAN - AN 11-YEAR STATEWIDE EXPERIENCE [J].
KATZ, DJ ;
STANLEY, JC ;
ZELENOCK, GB .
JOURNAL OF VASCULAR SURGERY, 1994, 19 (05) :804-817
[4]  
Moore W S, 1995, Cardiovasc Surg, V3, P109, DOI 10.1016/0967-2109(95)90882-6
[5]   Transfemoral endovascular repair of abdominal aortic aneurysm: Results of the North American EVT phase 1 trial [J].
Moore, WS ;
Rutherford, RB .
JOURNAL OF VASCULAR SURGERY, 1996, 23 (04) :543-553
[6]   Aortoiliac aneurysms: Management with endovascular stent-graft placement [J].
Murphy, KD ;
Richter, GM ;
Henry, M ;
Encarnacion, CE ;
Le, VA ;
Palmaz, JC .
RADIOLOGY, 1996, 198 (02) :473-480
[7]   ENDOVASCULAR REPAIR OF ABDOMINAL AORTIC-ANEURYSMS AND OTHER ARTERIAL LESIONS [J].
PARODI, JC .
JOURNAL OF VASCULAR SURGERY, 1995, 21 (04) :549-557
[8]  
ROZWNBLIT A, 1995, AM J ROENTGEROL, V165, P1473
[9]   COMPARISON OF THE UTILITY OF CT SCANS AND INTRAVASCULAR ULTRASOUND IN ENDOVASCULAR AORTIC GRAFTING [J].
VERBIN, C ;
SCOCCIANTI, M ;
KOPCHOK, G ;
DONAYRE, C ;
WHITE, RA .
ANNALS OF VASCULAR SURGERY, 1995, 9 (05) :434-440