Ultrasensitive near-infrared cavity ring-down spectrometer for precise line profile measurement

被引:24
作者
Gao, Bo [1 ]
Jiang, Wei [1 ]
Liu, An-Wen [1 ]
Lu, Yan [1 ]
Cheng, Cun-Feng [1 ]
Cheng, Guo-Sheng [1 ]
Hu, Shui-Ming [1 ]
机构
[1] Univ Sci & Technol China, Dept Chem Phys, Hefei Natl Lab Phys Sci Microscale, Hefei 230026, Peoples R China
关键词
Doppler broadening; Fabry-Perot interferometers; infrared spectra; infrared spectrometers; organic compounds; ring lasers; sapphire; spectrochemical analysis; titanium; ABSORPTION-MEASUREMENTS; MU-M; SPECTROSCOPY; BAND; (CH4)-C-12; RADICALS;
D O I
10.1063/1.3385675
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A cavity ring-down (CRD) spectrometer is built with a continuous-wave Ti:sapphire ring laser. Using a pair of R similar to 0.999 95 high-reflective mirrors, the noise-equivalent minimum detectable absorption loss reaches 7x10(-11)/cm over the spectral range of 780-830 nm. A thermal-stabilized Fabry-Perot interferometer is applied to calibrate the CRD spectrum with an accuracy of 1x10(-4) cm(-1). The quantitative measurement is carried out for the line profile measurements of some overtone absorption lines of C(2)H(2) near 787 nm. Doppler determined line shape has been observed with milli-Torr acetylene gas in the ring-down cavity. The instrumental line width is estimated from the line profile fitting to be < 1x10(-4) cm(-1). It demonstrates that the CRD spectrometer with extremely high sensitivity is also very suitable for quantitative measurements including precise line profile studies in the near-infrared.
引用
收藏
页数:5
相关论文
共 26 条
[1]   Broad-band cavity ring-down spectroscopy [J].
Ball, SM ;
Jones, RL .
CHEMICAL REVIEWS, 2003, 103 (12) :5239-5262
[2]   Cavity ring-down spectroscopy: Experimental schemes and applications [J].
Berden, G ;
Peeters, R ;
Meijer, G .
INTERNATIONAL REVIEWS IN PHYSICAL CHEMISTRY, 2000, 19 (04) :565-607
[3]   Absorption spectroscopy in high-finesse cavities for atmospheric studies [J].
Brown, SS .
CHEMICAL REVIEWS, 2003, 103 (12) :5219-5238
[4]  
BUSCH KW, 1999, ACS SYM SER, V720, P7
[5]   Direct determination of the Boltzmann constant by an optical method [J].
Daussy, C. ;
Guinet, M. ;
Amy-Klein, A. ;
Djerroud, K. ;
Hermier, Y. ;
Briaudeau, S. ;
Borde, Ch. J. ;
Chardonnet, C. .
PHYSICAL REVIEW LETTERS, 2007, 98 (25)
[6]   A frequency-stabilized difference frequency generation laser spectrometer for precise line profile studies in the midinfrared [J].
Deng, Wen-Ping ;
Gao, Bo ;
Cheng, Cun-Feng ;
Cheng, Guo-Sheng ;
Hu, Shui-Ming ;
Zhu, Qing-Shi .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (12)
[7]   A jet-cooled infrared spectrum of the formic acid dimer by cavity ring-down spectroscopy [J].
Ito, F ;
Nakanaga, T .
CHEMICAL PHYSICS LETTERS, 2000, 318 (06) :571-577
[8]   Very high sensitivity CW-cavity ring down spectroscopy:: Application to the a 1Δg(0)-X 3Σ-g(1) O2 band near 1.58 μm [J].
Kassi, S ;
Romanini, D ;
Campargue, A ;
Bussery-Honvault, B .
CHEMICAL PHYSICS LETTERS, 2005, 409 (4-6) :281-287
[9]   A compact broadband cavity enhanced absorption spectrometer for detection of atmospheric NO2 using light emitting diodes [J].
Langridge, Justin M. ;
Ball, Stephen M. ;
Jones, Roderic L. .
ANALYST, 2006, 131 (08) :916-922
[10]   High sensitivity CW-cavity ring down spectroscopy of water in the region of the 1.5 μm atmospheric window [J].
Macko, P ;
Romanini, D ;
Mikhailenko, SN ;
Naumenko, OV ;
Kassi, S ;
Jenouvrier, A ;
Tyuterev, VG ;
Campargue, A .
JOURNAL OF MOLECULAR SPECTROSCOPY, 2004, 227 (01) :90-108