Calculation of electric fields induced by body and head motion in high-field MRI

被引:55
作者
Liu, F [1 ]
Zhao, HW [1 ]
Crozier, S [1 ]
机构
[1] Univ Queensland, Ctr Magnet Resonance, Sch Informat Technol & Elect Engn, Brisbane, Qld 4072, Australia
基金
澳大利亚研究理事会;
关键词
MRI; high magnetic field; induced current; human body model;
D O I
10.1016/S1090-7807(02)00180-5
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
In modern magnetic resonance imaging (MRI), patients are exposed to strong, nonuniform static magnetic fields outside the central imaging region, in which the movement of the body may be able to induce electric currents in tissues which could be possibly harmful. This paper presents theoretical investigations into the spatial distribution of induced electric fields and currents in the patient when moving into the MRI scanner and also for head motion at various positions in the magnet. The numerical calculations are based on an efficient, quasi-static, finite-difference scheme and an anatomically realistic, full-body, male model. 3D field profiles from an actively shielded 4T magnet system are used and the body model projected through the field profile with a range of velocities. The simulation shows that it possible to induce electric fields/currents near the level of physiological significance under some circumstances and provides insight into the spatial characteristics of the induced fields. The results are extrapolated to very high field strengths and tabulated data shows the expected induced currents and fields with both movement velocity and field strength. (C) 2003 Elsevier Science (USA). All rights reserved.
引用
收藏
页码:99 / 107
页数:9
相关论文
共 27 条
[1]
[Anonymous], 1996, Handbook of biological effects of electromagnetic fields
[2]
THE ACTIVATING FUNCTION FOR MAGNETIC STIMULATION DERIVED FROM A 3-DIMENSIONAL VOLUME CONDUCTOR MODEL [J].
BASSER, PJ ;
WIJESINGHE, RS ;
ROTH, BJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (11) :1207-1210
[3]
BEISCHER DE, 1964, AEROSPACE MED, V35, P939
[4]
Bowtell R, 2000, MAGNET RESON MED, V44, P782, DOI 10.1002/1522-2594(200011)44:5<782::AID-MRM16>3.0.CO
[5]
2-7
[6]
BUDINGER TF, 1987, BOOK ABSTR, P183
[7]
The stochastic design of force-minimized compact magnets for high-field magnetic resonance imaging applications [J].
Crozier, S ;
Snape-Jenkinson, CJ ;
Forbes, LK .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (02) :4014-4022
[8]
Gandhi OP, 1999, MAGN RESON MED, V41, P816, DOI 10.1002/(SICI)1522-2594(199904)41:4<816::AID-MRM22>3.0.CO
[9]
2-5
[10]
Cognitive, cardiac, and physiological safety studies in ultra high field magnetic resonance imaging [J].
Kangarlu, A ;
Burgess, RE ;
Zhu, H ;
Nakayama, T ;
Hamlin, RL ;
Abduljalil, AM ;
Robitaille, PML .
MAGNETIC RESONANCE IMAGING, 1999, 17 (10) :1407-1416