The effect of camera geometry on singles flux, scatter fraction and trues and randoms sensitivity for cylindrical 3D PET - A simulation study.

被引:53
作者
Badawi, RD [1 ]
Kohlmyer, SG
Harrison, RL
Vannoy, SD
Lewellen, TK
机构
[1] Univ Washington, Div Nucl Med, Seattle, WA 98195 USA
[2] Dana Farber Canc Inst, Div Nucl Med, Boston, MA 02115 USA
[3] Harvard Univ, Sch Med, Boston, MA 02115 USA
关键词
D O I
10.1109/23.856575
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Preliminary results of an assessment of the effects of changing the axial field of view (AFOV) and detector ring diameter (DRD) of a cylindrical PET tomograph on count-rate performance are presented. The assessment was made using Monte Carlo simulations of an anthropomorphic phantom based on the Zubal phantom. This phantom was modified to include cylinders approximating arms and legs, and was sequentially stepped through the AFOV to simulate a whole-body scan covering an axial region of interest of Im. DRD was varied from similar to 60 cm to similar to 108 cm, and AFOV was varied from 10 cm to 60 cm. A simple activity distribution and deadtime model was assumed to allow the calculation of noise-equivalent count (NEC) rates for a situation similar to that of a typical 18F-FDG study. Both the scatter fraction and singles flux were found to be strongly dependent on DRD, but only weakly dependent on AFOV when the latter was greater than similar to 25 cm. Trues and randoms sensitivity were strongly dependent on AFOV, and randoms sensitivity was also strongly dependent on DRD. Scatter and singles flux do not appear to be limiting factors for extended AFOV configurations, and randoms rates, while high, appear to be manageable with existing detector technology. This initial assessment suggests that for whole-body applications, substantial gains in NEC may be possible by extending the AFOV.
引用
收藏
页码:1228 / 1232
页数:5
相关论文
共 13 条
[1]   Optimization of noise-equivalent count rates in 3D PET [J].
Badawi, RD ;
Marsden, PK ;
Cronin, BF ;
Sutcliffe, JL ;
Maisey, MN .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (09) :1755-1776
[2]   A SIMPLE DATA LOSS MODEL FOR POSITRON CAMERA SYSTEMS [J].
ERIKSSON, L ;
WIENHARD, K ;
DAHLBOM, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1566-1570
[3]  
HARRISON RL, 1993, C REC NUCL SCI S, V2, P1154
[4]  
JONES T, 1996, J NUCL MED, V35
[5]  
KOHLMYER SG, 1999, C REC IEEE NSS MIC 1
[6]  
MEIJA AA, 1991, J NUCL MED, V32, P699
[7]   A count rate model for PET and its application to an LSO HR PLUS scanner [J].
Moisan, C ;
Rogers, JG ;
Douglas, JL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1219-1224
[8]  
*NAT EL MAN ASS, 1994, NEMA STAND PUBL NU, V2
[9]   SIMULATION STUDIES FOR CYLINDRICAL POSITRON TOMOGRAPHY [J].
STEARNS, CW ;
BURNHAM, CA ;
CHESLER, DA ;
BROWNELL, GL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :708-711
[10]   MEASURING PET SCANNER SENSITIVITY - RELATING COUNTRATES TO IMAGE SIGNAL-TO-NOISE RATIOS USING NOISE EQUIVALENT COUNTS [J].
STROTHER, SC ;
CASEY, ME ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :783-788