Electrocatalytic detection of aurothiomalate on carbon electrodes -: Application as a non-enzymatic label to the quantification of proteins

被引:9
作者
de la Escosura-Muñiz, A [1 ]
González-García, MB [1 ]
Costa-García, A [1 ]
机构
[1] Univ Oviedo, Dept Quim Fis & Analit, Oviedo 33006, Asturias, Spain
关键词
ionic gold; electroactive label; electrocatalytic deposition of silver; aurothiomalate; glassy carbon electrode; bovine serum albumin; rabbit IgG;
D O I
10.1016/j.aca.2004.04.064
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new method to determine the concentration of protein solutions, bovine serum albumin and rabbit IgG anti human IgM (RIgG), has been developed in this work. This method is based on the catalytic effect of the electrodeposited gold on glassy carbon electrodes upon silver electrodeposition. Thus, ionic gold was used as an electroactive label of proteins. Proteins were labelled using a sodium aurothiomalate solution (gold(l) complex). Both solutions were prepared in unbuffered solutions of 0.15 M NaCl, pH 7.5, and the binding was carried out at 37 degreesC for 24 h. Then, free gold was separated from the labelled protein by a dialysis system, against a 0.15 M NaCl, pH 7.5, solution for 24 h at room temperature. The gold(I) bound to the protein, which was previously adsorbed on the electrode surface, was electrodeposited in 0.1 M HCl at -1.00 V for 5 min, and then oxidised in 0.1 M H2SO4 at a fixed potential, for 1 min. Then, silver electrodeposition at an adequate potential for 1 min followed by anodic stripping voltammetry were carried out in aqueous 1.0 M NH3-2.0 x 10(-4) M AgNO3. Thus, a good linear relationship between the anodic stripping peak and the concentration of labelled protein was found, achieving limits of detection of about 1 x 10(-8) M. The RIgG gold labelling was also used for monitoring an immunological reaction, against a biotinylated anti rabbit IgG which was immobilised on the electrode surface through the streptavidin:biotin interaction, showing that the labelling of RIgG with gold did not affect the antibody: antigen interaction. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:355 / 363
页数:9
相关论文
共 30 条
[1]  
ANVARHUSEIN A, 1981, J CHEM SOC DALTON, P1657
[2]  
COFFER MT, 1986, INORG CHEM, V25, P333
[3]   COMPARATIVE VOLTAMMETRIC STUDY OF 2,4-DINITROPHENOL (DNP), ALBUMIN AND DNP-ALBUMIN AS AN ANALYTICAL APPROACH TO THE USE OF DNP AS A UNIVERSAL LABEL IN IMMUNOELECTROCHEMICAL ASSAYS [J].
COSTAGARCIA, A ;
FERNANDEZABEDUL, MT ;
TUNONBLANCO, P .
TALANTA, 1994, 41 (07) :1191-1200
[4]   Disposable amperometric immunosensor system for rabbit IgG using a conducting polymer modified screen-printed electrode [J].
Darain, F ;
Park, SU ;
Shim, YB .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (5-6) :773-780
[5]  
DELAESCOSURAMUN.A, IN PRESS ELECTROANAL
[6]   An electrochemical metalloimmunoassay based on a colloidal gold label [J].
Dequaire, M ;
Degrand, C ;
Limoges, B .
ANALYTICAL CHEMISTRY, 2000, 72 (22) :5521-5528
[7]   HOMOGENEOUS FERROCENE-MEDIATED AMPEROMETRIC IMMUNOASSAY [J].
DIGLERIA, K ;
HILL, HAO ;
MCNEIL, CJ ;
GREEN, MJ .
ANALYTICAL CHEMISTRY, 1986, 58 (06) :1203-1205
[8]   HETEROGENEOUS IMMUNOASSAY FOR SERUM-PROTEINS BY DIFFERENTIAL PULSE ANODIC-STRIPPING VOLTAMMETRY [J].
DOYLE, MJ ;
HALSALL, HB ;
HEINEMAN, WR .
ANALYTICAL CHEMISTRY, 1982, 54 (13) :2318-2322
[9]   REACTIONS OF THE METALLODRUG AURANOFIN [(1-THIO-BETA-D-GLUCOPYRANOSE-2,3,4,6-TETRAACETATO-S)(TRIETHYLPHOSPHINE)GOLD] WITH BIOLOGICAL LIGANDS STUDIED BY RADIOISOTOPE METHODOLOGY [J].
ECKER, DJ ;
HEMPEL, JC ;
SUTTON, BM ;
KIRSCH, R ;
CROOKE, ST .
INORGANIC CHEMISTRY, 1986, 25 (18) :3139-3143
[10]  
Gonzalez Garcia M. B., 1995, Bioelectrochemistry and Bioenergetics, V38, P389, DOI 10.1016/0302-4598(95)01813-T