Lubberts effect in columnar phosphors

被引:54
作者
Badano, A
Gagne, RM
Gallas, BD
Jennings, RJ
Boswell, JS
Myers, KJ
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Lab Assessment Med Imaging Syst, Rockville, MD 20857 USA
[2] Natl Inst Biomed Imaging & Bioengn, NIH, Rockville, MD 20857 USA
关键词
noise transfer; phosphor blur; modulation transfer function; Monte Carlo simulation; columnar phosphor; Cesium Iodide;
D O I
10.1118/1.1796151
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Noise transfer in granular x-ray imaging phosphor screens is not proportional to the square of the magnitude of the signal transfer when the transfer properties are considered for the entire screen thickness, unless appropriately weighted at each depth of interaction. This property, known as the Lubberts effect, has not yet been studied in columnar structured screens because of a lack of a generalized description of the depth-dependent light transport. In this paper, we investigate the signal and noise transfer characteristics of columnar phosphors used in digital mammography detectors using DETECT-II, an optical Monte Carlo light transport simulation code. We first validate our choice of optical parameters for the description of granular and columnar screens using published normalized modulation transfer (MTF) experimental data. Our calculations of MTF match empirically measured MTFs for a granular film/screen analog system, and for an indirect x-ray digital imaging system with CsI:T1 screen representative of digital mammography systems. Using the depth-dependent spread functions and collection efficiencies, we calculate the signal and noise transfer functions and the Lubberts fraction, which is the ratio of the signal transfer function to the noise transfer function, for different screen thicknesses of granular and columnar phosphors. We find that the Lubberts fraction of a 85 mum granular screen model corresponding to a Gd2O2S:Tb screen is similar to the fraction for a 100 mum columnar CsI:T1 screen.
引用
收藏
页码:3122 / 3131
页数:10
相关论文
共 38 条
[1]   Image degradation by glare in radiologic display devices [J].
Badano, A ;
Flynn, MJ .
IMAGE DISPLAY - MEDICAL IMAGING 1997, 1997, 3031 :222-231
[2]   Optical blur and collection efficiency in columnar phosphors for X-ray imaging [J].
Badano, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 508 (03) :467-479
[3]   Monte Carlo analysis of the spectral photon emission and extraction efficiency of organic light-emitting devices [J].
Badano, A ;
Kanicki, J .
JOURNAL OF APPLIED PHYSICS, 2001, 90 (04) :1827-1830
[4]  
BADANO A, 1999, THESIS U MICHIGAN
[5]  
Barrett H., 1996, RADIOLOGICAL IMAGING
[6]   PHOTOGRAPHIC EMULSION LAYER AS A 3-DIMENSIONAL RECORDING MEDIUM [J].
BERG, WF .
APPLIED OPTICS, 1969, 8 (12) :2407-&
[7]  
BERGER MJ, 1982, NBSIR822550 NAT BUR
[8]  
BERGER MJ, XCOM PHOTON CROSS DA
[9]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[10]  
BORN M, 1965, PRINCIPLES OPTICS