Fixed-energy biphasic waveform defibrillation in a pediatric model of cardiac arrest and resuscitation

被引:44
作者
Tang, WC
Weil, MH
Jorgenson, D
Klouche, K
Morgan, C
Yu, T
Sun, SJ
Snyder, D
机构
[1] Inst Crit Care Med, Palm Springs, CA 92262 USA
[2] Univ So Calif, Keck Sch Med, Los Angeles, CA USA
[3] Philips Med Syst Heartstream, Seattle, WA USA
关键词
cardiopulmonary resuscitation; pediatric cardiac arrest; ventricular fibrillation; defibrillation; postresuscitation myocardial dysfunction; piglet cardiopulmonary resuscitation;
D O I
10.1097/00003246-200212000-00019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: For adults, 150-J fixed-energy, impedance-compensating biphasic truncated exponential (ICBTE) shocks are now effectively used in automated defibrillators. However, the high energy levels delivered by adult automated defibrillators preclude their use for pediatric patients. Accordingly, we investigated a method by which adult automated defibrillators may be adapted to deliver a 50-J ICBTE shock for pediatric defibrillation. Design: Prospective, randomized study. Setting: A university-affiliated research institution. Subject: Domestic piglets. Interventions: We initially investigated four groups of anesthetized mechanically ventilated piglets weighing 3.8, 7.5,15, and 25 kg. Ventricular fibrillation was induced with an AC current delivered to the right ventricular endocardium. After 7 mins of untreated ventricular fibrillation, a conventional manual defibrillator was used to deliver up to three 50-J ICBTE shocks. If ventricular fibrillation was not reversed, a 1-min interval of precordial compression preceded a second sequence of up to three shocks. The protocol was repeated until spontaneous circulation was restored, or for a total of 15 mins. In a second set of experiments, we evaluated a 150-J biphasic adult automated defibrillator that was operated in conjunction with energy-reducing electrodes such as to deliver 50-J shocks. The same resuscitation protocol was then exercised on piglets weighing 3.7, 13.5, and 24.2 kg. Measurements and Main Results: All animals were successfully resuscitated. Postresuscitation hemodynamic and myocardial function quickly returned to baseline values in both experimental groups, and all animals survived. Conclusion: An adaptation of a 150-J biphasic adult automated defibrillator in which energy-reducing electrodes delivered 50-J shocks successfully resuscitated animals ranging from 3.7 to 25 kg without compromise of postresuscitation myocardial function or survival.
引用
收藏
页码:2736 / 2741
页数:6
相关论文
共 28 条
[1]   Accurate recognition and effective treatment of ventricular fibrillation by automated external defibrillators in adolescents [J].
Atkins, DL ;
Hartley, LL ;
York, DK .
PEDIATRICS, 1998, 101 (03) :393-397
[2]  
Clark CB, 1999, CIRCULATION, V100, P90
[3]   SURVIVAL TO LEAVE HOSPITAL FROM VENTRICULAR-FIBRILLATION [J].
DUNN, HM ;
MCCOMB, JM ;
MACKENZIE, G ;
ADGEY, AAJ .
AMERICAN HEART JOURNAL, 1986, 112 (04) :745-751
[4]  
EISENBERG M, 1983, ANN EMERG MED, V12, P672, DOI 10.1016/S0196-0644(83)80413-2
[5]   MYOCARDIAL DAMAGE FOLLOWING TRANS-THORACIC DIRECT-CURRENT COUNTERSHOCK IN NEWBORN PIGLETS [J].
GABA, DM ;
TALNER, NS .
PEDIATRIC CARDIOLOGY, 1982, 2 (04) :281-288
[6]  
Gliner Bradford E., 1998, Biomedical Instrumentation and Technology, V32, P631
[7]  
Institute of Laboratory Animal Resources, 1985, NIH PUBL
[8]   DECREASED DEFIBRILLATOR-INDUCED DYSFUNCTION WITH BIPHASIC RECTANGULAR WAVEFORMS [J].
JONES, JL ;
JONES, RE .
AMERICAN JOURNAL OF PHYSIOLOGY, 1984, 247 (05) :H792-H796
[9]   DYSFUNCTION AND SAFETY FACTOR STRENGTH-DURATION CURVES FOR BIPHASIC DEFIBRILLATOR WAVE-FORMS [J].
JONES, JL ;
MILNE, KB .
AMERICAN JOURNAL OF PHYSIOLOGY, 1994, 266 (01) :H263-H271
[10]   ULTRASTRUCTURAL INJURY TO CHICK MYOCARDIAL-CELLS INVITRO FOLLOWING ELECTRIC COUNTERSHOCK [J].
JONES, JL ;
PROSKAUER, CC ;
PAULL, WK ;
LEPESCHKIN, E ;
JONES, RE .
CIRCULATION RESEARCH, 1980, 46 (03) :387-394