New fast screening method for organochlorine pesticides in water by using solid-phase microextraction with fast gas chromatography and a pulsed-discharge electron capture detector

被引:38
作者
Jackson, GP [1 ]
Andrews, ARJ [1 ]
机构
[1] Ohio Univ, Dept Chem, Ctr Intelligent Chem Instrumentat, Athens, OH 45701 USA
关键词
solid-phase microextraction; organochlorine pesticides; fast gas chromatography; pulsed-discharge electron capture detector; carbon dioxide cryotrap;
D O I
10.1039/a706407a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new fast screening method has been developed for the analysis of organochlorine pesticides in water that takes less than 10 min to perform and has detection limits of the order of 10 ng l(-1), Pesticide extraction was achieved by solid-phase microextraction and the separation was performed using a micro-bore (0.1 mm) capillary column, A laboratory-built cryotrap system was used to focus the analytes prior to introduction to the column and a pulsed-discharge electron capture detector enabled sensitive, selective measurements of the pesticide peaks to be made. The 100 mu m polydimethylsilicone (PDMS) coated fibre gave better recoveries than the 30 or 7 mu m film PDMS fibres and an inter-fibre study comparing three 100 mu m film fibres showed good reproducibility, With extraction times of 2 min, the procedure was found to be linear over the range 0.01-1.2 ng ml(-1). The procedure was tested with a real river water sample.
引用
收藏
页码:1085 / 1090
页数:6
相关论文
共 25 条
[1]   Solid phase microextraction for quantitative analysis in nonequilibrium situations [J].
Ai, J .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :1230-1236
[2]  
[Anonymous], 1996, ANALYST, DOI DOI 10.1039/an9962100929
[3]  
[Anonymous], 1991, EPA600488039
[4]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[5]   ENVIRONMENTAL-PROTECTION-AGENCY AND OTHER METHODS FOR THE DETERMINATION OF PRIORITY PESTICIDES AND THEIR TRANSFORMATION PRODUCTS IN WATER [J].
BARCELO, D .
JOURNAL OF CHROMATOGRAPHY, 1993, 643 (1-2) :117-143
[6]   Determination of acidic chlorinated pesticides in water: Comparison of official EPA method 515.1 and liquid-solid extraction HPLC/UV and HPLC/PBMS analyses [J].
Bruner, F ;
Berloni, A ;
Palma, P .
CHROMATOGRAPHIA, 1996, 43 (5-6) :279-284
[7]   Characterization of the fussed discharge electron capture detector [J].
Cai, HM ;
Wentworth, WE ;
Stearns, SD .
ANALYTICAL CHEMISTRY, 1996, 68 (07) :1233-1244
[8]  
DESTY DH, 1965, ADV CHROMATOGR, V1, P199
[9]   ELECTRICALLY HEATED COLD TRAP INLET SYSTEM FOR HIGH-SPEED GAS-CHROMATOGRAPHY [J].
EWELS, BA ;
SACKS, RD .
ANALYTICAL CHEMISTRY, 1985, 57 (14) :2774-2779
[10]   SAMPLE INTRODUCTION APPROACHES FOR SOLID-PHASE MICROEXTRACTION RAPID GC [J].
GORECKI, T ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3265-3274