ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart

被引:170
作者
Kachelriess, M [1 ]
Ulzheimer, S [1 ]
Kalender, WA [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
关键词
computed tomography (CT); multi-slice spiral CT; heart; 4D reconstruction;
D O I
10.1118/1.1286552
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Subsecond spiral computed tomography (CT) offers great potential for improving heart imaging. The new multi-row detector technology adds significantly to this potential. We therefore developed and validated dedicated cardiac reconstruction algorithms for imaging the heart with subsecond multi-slice spiral CT utilizing electrocardiogram (ECG) information. The single-slice cardiac z-interpolation algorithms 180 degrees CI and 180 degrees CD [Med. Phys. 25, 2417-2431 (1998)] were generalized to allow imaging of the heart for M-slice scanners. Two classes of algorithms were investigated: 180 degrees MCD (multi-slice cardio delta), a partial scan reconstruction of 180 degrees+ delta data with delta <Phi (fan angle) resulting in effective scan times of 250 ms (central ray) when a 0.5 a rotation mode is available, and 180 degrees MCI (multi-slice cardio interpolation), a piecewise weighted interpolation between successive spiral data segments belonging to the same heart phase, potentially providing a relative temporal resolution of 12.5% of the heart cycle when a four-slice scanner is used and the table increment is chosen to be greater than or equal to the collimated slice thickness. Data segments are selected by correlation with the simultaneously recorded ECG signal. Theoretical studies, computer simulations, as well as patient measurements were carried out for a multi-slice scanner providing M=4 slices to evaluate these new approaches and determine the optimal scan protocol. Both algorithms, 180 degrees MCD and 180 degrees MCI, provide significant improvements in image quality, including extremely arythmic cases. Artifacts in the reconstructed images as well as in 3D displays such as multiplanar reformations were largely reduced as compared to the standard z-interpolation algorithm 180 degrees MLI (multi-slice linear interpolation). Image quality appears adequate for precise calcium scoring and CT angiography of the coronary arteries with conventional subsecond multislice spiral CT. It turned out that for heart rates f(H)greater than or equal to 70 min(-1) the partial scan approach 180 degrees MCD yields unsatisfactory results as compared to 180 degrees MCI. Our theoretical considerations show that a freely selectable scanner rotation time chosen as a function of the patient's heart rate, would further improve the relative temporal resolution and thus further reduce motion artifacts. In our case an additional 0.6 s mode besides the available 0.5 s mode would be very helpful. Moreover, if technically feasible, lower rotation times such as 0.3 s or even less would result in improved image quality. The use of multi-slice techniques for cardiac CT together with the new z-interpolation methods improves the quality of heart imaging significantly. The high temporal resolution of 180 degrees MCI is adequate for spatial and temporal tracking of anatomic structures of the heart (4D reconstruction). (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00208-X].
引用
收藏
页码:1881 / 1902
页数:22
相关论文
共 27 条
[1]   Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary-artery stenoses and occlusions [J].
Achenbach, S ;
Moshage, W ;
Ropers, D ;
Nossen, J ;
Daniel, WG .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (27) :1964-1971
[2]   QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[3]  
[Anonymous], 2000, COMPUTED TOMOGRAPHY
[4]   Elastic spline models for human cardiac motion estimation [J].
Chen, F ;
Suter, D .
IEEE NONRIGID AND ARTICULATED MOTION WORKSHOP, PROCEEDINGS, 1997, :120-127
[5]  
Hiraoka M, 1999, RADIOLOGY, V213P, P401
[6]  
Hsieh J, 1999, RADIOLOGY, V213P, P401
[7]   Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[8]  
KACHELRIES M, 1998, ADV CT, V4, P137
[9]  
Kachelriess M, 1998, RADIOLOGY, V209P, P323
[10]   Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart [J].
Kachelriess, M ;
Kalender, WA .
MEDICAL PHYSICS, 1998, 25 (12) :2417-2431