Performance characteristics of six third-generation assays for thyroid-stimulating hormone

被引:84
作者
Rawlins, ML
Roberts, WL
机构
[1] ARUP Inst Clin & Expt Pathol, Salt Lake City, UT USA
[2] Univ Utah, Hlth Sci Ctr, Dept Pathol, Salt Lake City, UT 84132 USA
关键词
D O I
10.1373/clinchem.2004.039156
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Thyroid-stimulating hormone (TSH) is used to detect primary hypo- and hyperthyroidism. Current guidelines for TSH assays recommend a functional sensitivity of less than or equal to 0.02 mIU/L. The protocol for determining the functional sensitivity of TSH assays specifies analyses of serum samples with two reagent lots over a 6- to 8-week period. Methods: We determined the functional sensitivities of the Access 2, ADVIA Centaur, ARCHITECT i2000, E170, IMMULITE 2000, and Vitros EG automated methods, using seven serum pools and two reagent lots for each method. Results: The observed functional sensitivities were as follows: Access 2, <0.020 mIU/L; ADVIA Centaur, 0.039 mIU/L; ARCHITECT i2000, <0.005 mIU/L; Elecsys E170, 0.011 mIU/L; IMMULITE 2000, 0.014 mIU/L; Vitros EG, 0.004 mIU/L. However, there were large differences between some method means for the seven serum pools. For the pool with the lowest TSH concentration, mean results were as follows: Access 2, 0.0203 mIU/L; ADVIA Centaur, 0.0085 mIU/L; ARCHITECT i2000, 0.0049 mIU/L; E170, 0.0098 mIU/L; IMMULITE 2000, 0.0077 mIU/L; Vitros ECi, 0.0014 mIU/L. Method-comparison studies using samples with TSH concentrations >0.2 mIU/L also showed method differences. The ARCHITECT i2000 method was the most precise at low TSH concentrations. Conclusions: TSH methods do not provide comparable results for serum pools with TSH concentrations <0.2 mIU/L or for patient results across the analytic measure- ment range. Further investigation into the cause of these differences and additional harmonization efforts are required. (C) 2004 American Association for Clinical Chemistry.
引用
收藏
页码:2338 / 2344
页数:7
相关论文
共 13 条
[1]  
[Anonymous], 2003, THYROID, V13, P3
[2]   BIOLOGICAL AND IMMUNOCHEMICAL CHARACTERIZATION OF RECOMBINANT HUMAN THYROTROPIN [J].
CANONNE, C ;
PAPANDREOU, MJ ;
MEDRI, G ;
VERRIER, B ;
RONIN, C .
GLYCOBIOLOGY, 1995, 5 (05) :473-481
[3]  
dHerbomez M, 1997, EUR J CLIN CHEM CLIN, V35, P609
[4]   A new modular chemiluminescence immunoassay analyser evaluated [J].
Ognibene, A ;
Drake, CJ ;
Jeng, KYS ;
Pascucci, TE ;
Hsu, S ;
Luceri, F ;
Messeri, G .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (03) :251-260
[5]  
Price A, 2001, CLIN CHEM, V47, P2067
[6]   A comparison of different sample matrices for evaluating functional sensitivity, imprecision and dilution linearity of the abbott ARCHITECT® i2000® TSH assay [J].
Quinn, FA ;
Scopp, R ;
Lach, A ;
Drake, C ;
Mo, M ;
Albright, J ;
Trimpe, K .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (07) :709-712
[7]   Discrepancies between thyrotropin (TSH) measurement by four sensitive immunometric assays [J].
Rasmussen, AK ;
Hilsted, L ;
Perrild, H ;
Christiansen, E ;
SiersbaekNielsen, K ;
FeldtRasmussen, U .
CLINICA CHIMICA ACTA, 1997, 259 (1-2) :117-128
[8]   Utility of third generation thyrotropin assays in thyroid function testing [J].
Saller, B ;
Broda, N ;
Heydarian, R ;
Görges, R ;
Mann, K .
EXPERIMENTAL AND CLINICAL ENDOCRINOLOGY & DIABETES, 1998, 106 :S29-S33
[9]  
Sapin R, 2000, CLIN CHEM, V46, P418
[10]  
Saw S, 1999, CLIN CHEM, V45, P578