Vertical focusing device utilizing dielectrophoretic force and its application on microflow cytometer

被引:82
作者
Lin, CH [1 ]
Lee, GB
Fu, LM
Hwey, BH
机构
[1] Natl Sun Yat Sen Univ, Dept Mech & Electromech Engn, Kaohsiung 804, Taiwan
[2] Natl Cheng Kung Univ, Dept Engn Sci, Tainan 701, Taiwan
[3] Natl Pingtung Univ Sci & Technol, Grad Inst Mat Engn, Pingtung 912, Taiwan
关键词
dielectrophoretic force; flow cytometer; hydrodynamic force; microelectromechanical systems (MEMS); three-dimensional (3-D) focusing;
D O I
10.1109/JMEMS.2004.838352
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Focusing of particles/cells in the vertical direction inside a micromachined flow cytometer is a critical issue while using an embedded optical detection system aligned with microchannels. Even if the particles/cells have been focused centrally in the horizontal direction using coplanar sheath flows, appreciable errors may still arise if they are randomly distributed in the vertical direction. This paper presents a vertical focusing device utilizing dielectrophoretic (DEP) forces and its application on micromachined flow cytometer. A pair of parallel microelectrodes is deposited on the upper and bottom surface of the microfluidic channel to drive particles/cells into the vertical center of the sample flow. This new microfluidic device is capable of three-dimensional (3-D) focusing of microparticles/cells and thus improves the uniformity of the optical detection signals. This 3-D focusing feature of the sample now is realized utilizing the combination of dielectrophoretic and hydrodynamic forces. Initially, two sheath flows are used to focus the sample flow horizontally by means of hydrodynamic forces, and then two embedded planar electrodes apply negative DEP forces to focus the particles/cells vertically. A new micromachined flow cytometer integrated with an embedded optical detection mechanism is then demonstrated. Numerical simulation is used to analyze the operation conditions and the dimension of the microelectrodes for DEP manipulation. The dynamic trace of the moving particles/cells within a flow stream under the DEP manipulation is calculated numerically. Micro polystyrene beads and diluted human red blood cells (RBC) are used to test the performance of the propose device. The experimental results confirm the suitability of the proposed device for applications requiring precise counting of particles or cells. Experimental data indicate the proposed method can provide more stable signals over the other types of micromachined flow cytometers that were previously reported.
引用
收藏
页码:923 / 932
页数:10
相关论文
共 28 条
[1]  
BARDELL RL, 2001, P SPIE BIOM OPT S BI, V4265
[2]   NEW FLOW CYTOMETRIC CAPABILITIES AT THE NATIONAL FLOW-CYTOMETRY RESOURCE [J].
CRAM, LS ;
MARTIN, JC ;
STEINKAMP, JA ;
YOSHIDA, TM ;
BUICAN, TN ;
MARRONE, BL ;
JETT, JH ;
SALZMAN, G ;
SKLAR, L .
PROCEEDINGS OF THE IEEE, 1992, 80 (06) :912-917
[3]   Design and fabrication of travelling wave dielectrophoresis structures [J].
Cui, L ;
Morgan, H .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2000, 10 (01) :72-79
[4]   Optical particle detection integrated in a dielectrophoretic lab-on-a-chip [J].
Cui, L ;
Zhang, T ;
Morgan, H .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2002, 12 (01) :7-12
[5]   Manipulation of microparticles using new modes of traveling-wave-dielectrophoretic forces: Numerical simulation and experiments [J].
Fu, LM ;
Lee, GB ;
Lin, YH ;
Yang, RJ .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2004, 9 (02) :377-383
[6]   Electrokinetically driven micro flow cytometers with integrated fiber optics for on-line cell/particle detection [J].
Fu, LM ;
Yang, RJ ;
Lin, CH ;
Pan, YJ ;
Lee, GB .
ANALYTICA CHIMICA ACTA, 2004, 507 (01) :163-169
[7]   Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing [J].
Gawad, S ;
Schild, L ;
Renaud, P .
LAB ON A CHIP, 2001, 1 (01) :76-82
[8]   The application of flow cytometry to histocompatibility testing [J].
Horsburgh, T ;
Martin, S ;
Robson, AJ .
TRANSPLANT IMMUNOLOGY, 2000, 8 (01) :3-15
[9]   ELECTROKINETIC BEHAVIOR OF COLLOIDAL PARTICLES IN TRAVELING ELECTRIC-FIELDS - STUDIES USING YEAST-CELLS [J].
HUANG, Y ;
WANG, XB ;
TAME, JA ;
PETHIG, R .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1993, 26 (09) :1528-1535
[10]   Measurement of bacterial flagellar thrust by negative dielectrophoresis [J].
Hughes, MP ;
Morgan, H .
BIOTECHNOLOGY PROGRESS, 1999, 15 (02) :245-249