Measuring the local pressure amplitude in microchannel acoustophoresis

被引:214
作者
Barnkob, Rune [1 ]
Augustsson, Per [2 ]
Laurell, Thomas [2 ]
Bruus, Henrik [1 ]
机构
[1] Tech Univ Denmark, Dept Micro & Nanotechnol, DK-2800 Lyngby, Denmark
[2] Lund Univ, Dept Elect Measurements, S-22100 Lund, Sweden
基金
瑞典研究理事会;
关键词
ULTRASOUND STANDING-WAVE; MICROFLUIDIC CHANNELS; SUSPENDED PARTICLES; ACOUSTIC RESONANCES; LAYERED RESONATORS; CELL; SEPARATION; CHIP; FLOW; MICROPARTICLES;
D O I
10.1039/b920376a
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new method is reported on how to measure the local pressure amplitude and the Q factor of ultrasound resonances in microfluidic chips designed for acoustophoresis of particle suspensions. The method relies on tracking individual polystyrene tracer microbeads in straight water-filled silicon/glass microchannels. The system is actuated by a PZT piezo transducer attached beneath the chip and driven by an applied ac voltage near its eigenfrequency of 2 MHz. For a given frequency a number of particle tracks are recorded by a CCD camera and fitted to a theoretical expression for the acoustophoretic motion of the microbeads. From the curve fits we obtain the acoustic energy density, and hence the pressure amplitude as well as the acoustophoretic force. By plotting the obtained energy densities as a function of applied frequency, we obtain Lorentzian line shapes, from which the resonance frequency and the Q factor for each resonance peak are derived. Typical measurements yield acoustic energy densities of the order of 10 J/m(3), pressure amplitudes of 0.2 MPa, and Q factors around 500. The observed half wavelength of the transverse acoustic pressure wave is equal within 2% to the measured width w = 377 mu m of the channel.
引用
收藏
页码:563 / 570
页数:8
相关论文
共 45 条
[1]   ACOUSTIC RADIATION POTENTIAL ON A SPHERE IN PLANE, CYLINDRICAL, AND SPHERICAL STANDING WAVE FIELDS [J].
BARMATZ, M ;
COLLAS, P .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1985, 77 (03) :928-945
[2]  
Barnkob R., 2009, P M AC, V6, DOI [10.1121/1.3186746, DOI 10.1121/1.3186746]
[3]   Physical enviroment of 2-D animal cell aggregates formed in a short pathlength ultrasound standing wave trap [J].
Bazou, D ;
Kuznetsova, LA ;
Coakley, WT .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (03) :423-430
[4]   Ultrasonic agitation in microchannels [J].
Bengtsson, M ;
Laurell, T .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (07) :1716-1721
[5]  
BROWN D, 2009, TRACKER 2 60 FREE VI
[6]  
Bruus H., 2007, Theoretical microfluidics
[7]   Noninvasive acoustic cell trapping in a microfluidic perfusion system for online bioassays [J].
Evander, Mikael ;
Johansson, Linda ;
Lilliehorn, Tobias ;
Piskur, Jure ;
Lindvall, Magnus ;
Johansson, Stefan ;
Almqvist, Monica ;
Laurell, Thomas ;
Nilsson, Johan .
ANALYTICAL CHEMISTRY, 2007, 79 (07) :2984-2991
[8]  
Gor'kov L., 1962, SOV PHYS DOKL, V6, P773, DOI DOI 10.1039/C004504G
[9]   Harmonic Microchip Acoustophoresis: A Route to Online Raw Milk Sample Precondition in Protein and Lipid Content Quality Control [J].
Grenvall, Carl ;
Augustsson, Per ;
Folkenberg, Jacob Riis ;
Laurell, Thomas .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :6195-6200
[10]  
Groschl H, 1998, ACUSTICA, V84, P815