Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer

被引:136
作者
van Popele, NM
Bos, WJW
de Beer, NAM
van der Kuip, DAM
Hofman, A
Grobbee, DE
Witteman, JCM
机构
[1] Erasmus Univ, Med Ctr, Dept Epidemiol & Biostat, NL-3000 DR Rotterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Internal Med, NL-1105 AZ Amsterdam, Netherlands
[3] Univ Amsterdam, Acad Med Ctr, TNO Biomed Instrumentat, NL-1105 AZ Amsterdam, Netherlands
[4] Univ Utrecht, Med Ctr, Julius Ctr Patient Oriented Res, NL-3508 TC Utrecht, Netherlands
关键词
blood pressure monitoring; oscillometry; sphygmomanometry; arterial stiffness;
D O I
10.1161/01.HYP.36.4.484
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Oscillometric blood pressure devices tend to overestimate systolic blood pressure and underestimate diastolic blood pressure compared with sphygmomanometers. Recent studies indicate that discrepancies in performance between these devices may differ between healthy and diabetic subjects. Arterial stiffness in diabetics could be the underlying factor explaining these differences. We studied differences between a Dinamap oscillometric blood pressure monitor and a random-zero sphygmomanometer in relation to arterial stiffness in 1808 healthy elderly subjects. The study was conducted within the Rotterdam Study, a population-based cohort study of subjects aged 55 years and older. Systolic and diastolic blood pressure differences between a Dinamap and a random-zero sphygmomanometer were related to arterial stiffness, as measured by carotid-femoral pulse wave velocity. Increased arterial stiffness was associated with higher systolic and diastolic brood pressure readings by the Dinamap compared with the random-zero sphygmomanometer, independent of age, gender, and average mean blood pressure level of both devices. The beta -coefficient (95% CI) was 0.25 (0.00 to 0.50) mm Hg/(m/s) for the systolic blood pressure difference and 0.35 (0.20 to 0.50) mm Hg/(m/s) for the diastolic blood pressure difference. The results indicate that a Dinamap oscillometric blood pressure device, in comparison to a random-zero sphygmomanometer, overestimates systolic and diastolic blood pressure readings in subjects with stiff arteries.
引用
收藏
页码:484 / 488
页数:5
相关论文
共 18 条
[1]   ASSESSMENT OF ARTERIAL DISTENSIBILITY BY AUTOMATIC PULSE-WAVE VELOCITY-MEASUREMENT - VALIDATION AND CLINICAL-APPLICATION STUDIES [J].
ASMAR, R ;
BENETOS, A ;
TOPOUCHIAN, J ;
LAURENT, P ;
PANNIER, B ;
BRISAC, AM ;
TARGET, R ;
LEVY, BI .
HYPERTENSION, 1995, 26 (03) :485-490
[2]   Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ values [J].
Brown, WCB ;
Kennedy, S ;
Inglis, GC ;
Murray, LS ;
Lever, AF .
JOURNAL OF HUMAN HYPERTENSION, 1997, 11 (02) :75-93
[3]  
Duprez DA, 1996, J HUM HYPERTENS, V10, P251
[4]  
GOONASEKERA CDA, 1995, J HUM HYPERTENS, V9, P885
[5]   COMPARISON OF AUTOMATIC OSCILLOMETRIC ARTERIAL-PRESSURE MEASUREMENT WITH CONVENTIONAL AUSCULTATORY MEASUREMENT IN THE LABOR WARD [J].
HASAN, MA ;
THOMAS, TA ;
PRYSROBERTS, C .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (02) :141-144
[6]   DETERMINANTS OF DISEASE AND DISABILITY IN THE ELDERLY - THE ROTTERDAM ELDERLY STUDY [J].
HOFMAN, A ;
GROBBEE, DE ;
DEJONG, PTVM ;
VANDENOUWELAND, FA .
EUROPEAN JOURNAL OF EPIDEMIOLOGY, 1991, 7 (04) :403-422
[7]   ARTERIAL-WALL COMPLIANCE IN DIABETES [J].
LEHMANN, ED ;
GOSLING, RG ;
SONKSEN, PH .
DIABETIC MEDICINE, 1992, 9 (02) :114-119
[9]   Carotid arterial stiffness as a surrogate for aortic stiffness: Relationship between carotid artery pressure-strain elastic modulus and aortic pulse wave velocity [J].
Nagai, Y ;
Fleg, JL ;
Kemper, MK ;
Rywik, TM ;
Earley, CJ ;
Metter, EJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (02) :181-188
[10]   Does it matter whether blood pressure measurements are taken with subjects sitting or supine? [J].
Netea, RT ;
Smits, P ;
Lenders, JWM ;
Thien, T .
JOURNAL OF HYPERTENSION, 1998, 16 (03) :263-268