Evaluation of the Lactate Pro blood lactate analyser

被引:250
作者
Pyne, DB
Boston, T
Martin, DT
Logan, A
机构
[1] Australian Inst Sport, Dept Physiol & Appl Nutr, Belconnen, ACT 2616, Australia
[2] Univ Canberra, GADI Res Ctr, Canberra, ACT, Australia
[3] Australian Swimming Inc, Canberra, ACT, Australia
关键词
lactate; lactate analysis; reliability; accuracy; level of agreement;
D O I
10.1007/s004210050659
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
An evaluation of the hand-held portable Lactate Pro Analyser (KDK) was undertaken to assess its accuracy, reliability and versatility. Capillary blood samples were drawn from elite athletes in both laboratory and field settings and analysed in parallel. Accuracy was determined in relation to three other lactate analysers: (1) the ABL 700 Series Acid-Base analyser (n = 172 cases), (2) the Accusport Lactate Meter (n = 118 cases), and (3) the YSI 2300 Stat lactate analyser (n = 22 cases). The level of agreement was determined over the range of 1-18 mM. The repeatability of results between two different Lactate Pro analysers was also determined over the same range. Versatility was assessed in the field, where the Lactate Pro was used with elite athletes under a range of outdoor and indoor testing conditions. The correlations between the Lactate Pro and the ABL 700 Series Acid-Base analyser, YSI 2300 and Accusport were r = 0.98, r = 0.99, r = 0.97. The correlation between the two Lactate Pro analysers on the same sample (n = 96 cases) was r = 0.99. The level of agreement between the Lactate Pro and other analysers was generally less than +/- 2.0 mM over the physiological range of 1.0-18.0 mM (range of mean difference: -0.06 mM to 0.52 mM). The Lactate Pro was easy to operate and successfully completed the sample analysis in 100% of the tests performed. In summary, the Lactate Pro is accurate, reliable and exhibits a high degree of agreement with other lactate analysers.
引用
收藏
页码:112 / 116
页数:5
相关论文
共 17 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]  
[Anonymous], 1984, Journal of Swimming Research
[3]   Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine [J].
Atkinson, G ;
Nevill, AM .
SPORTS MEDICINE, 1998, 26 (04) :217-238
[4]  
CLOUGH M, 1997, SPORTS MED AUSTR CAN, P82
[5]  
Coyle E F, 1995, Exerc Sport Sci Rev, V23, P25
[6]   Evaluation of the Accusport® Lactate Analyzer [J].
Fell, JW ;
Rayfield, JM ;
Gulbin, JP ;
Gaffney, PT .
INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1998, 19 (03) :199-204
[7]   LACTATE CONCENTRATION DIFFERENCES IN PLASMA, WHOLE-BLOOD, CAPILLARY FINGER BLOOD AND ERYTHROCYTES DURING SUBMAXIMAL GRADED-EXERCISE IN HUMANS [J].
FOXDAL, P ;
SJODIN, B ;
RUDSTAM, H ;
OSTMAN, C ;
OSTMAN, B ;
HEDENSTIERNA, GC .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 1990, 61 (3-4) :218-222
[8]   THE EFFECT OF DIFFERENT BLOOD-SAMPLING SITES AND ANALYSES ON THE RELATIONSHIP BETWEEN EXERCISE INTENSITY AND 4.0 MMOL.L-1 BLOOD LACTATE CONCENTRATION [J].
FOXDAL, P ;
SJODIN, A ;
OSTMAN, B ;
SJODIN, B .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1991, 63 (01) :52-54
[9]  
Lowensteyn I., 1994, J SWIMMING RES, V10, P31
[10]  
MADSEN O, 1987, SWIMMING TECHN, V24, P21